21.6 Configuring D/A Output Items (Option)
This section explains the following settings for D/A output. This feature is available on models with the
/DA option.
• Measurement function
• Element and wiring unit
• Harmonic order
• D/A output range
Range mode, range maximum, and range minimum
►
“D/A Output (D/A Output Items; option)” in the features guide
Configuring D/A Output Items
Press
UTILITY
and then the
D/A Output Items
soft key to display the following screen.
Set the element and wiring unit (Element 1 to Element 6, ΣA to ΣC).
Set the measurement function
(None, other functions—for details on the various measurement functions, see
“Items That This Instrument Can Measure” in the features guide).
Output item
This display changes according to the Function, Element/Σ, and Order settings.
Set the harmonic order (Total, 0 to 500; /G5 or /G6 option).
You can set this setting when the measurement function includes a
harmonic order.
Select the mode of the D/A
output range (Fix, Manual).
Set the maximum and minimum
values of the range
(–9.999 T to 9.999 T).
These settings can be set when
Range Mode is set to Manual.
D/A output signal name
For details on the connector pinout and the D/A output signal assignment, see section 4.6 in the
getting started guide, IM WT1801E-03EN.