20-3

IM WT18O1E-02EN

20.2  Configuring TCP/IP Settings

This section explains the following TCP/IP settings for connecting this instrument to a network:
•  DHCP
  IP address, subnet mask, and default gateway
•  DNS
  Domain name, DNS server IP address, and domain suffix

 “TCP/IP (TCP/IP)” in the features guide

Configuring TCP/IP Settings (TCP/IP)

Press 

UTILITY

, the 

Network

 soft key, and then the 

TCP/IP

 soft key to display the following screen.

  

Set the DHCP (OFF, ON). 

Set these when DHCP is set to OFF.

•  IP address

•  Subnet mask

•  Default gateway

These are displayed when DNS is set to ON or Auto.

•  Domain name

•  DNS server IP address

  (1: primary, 2: secondary)

•  Domain suffix

  (1: primary, 2: secondary)
Applies the settings

Set the DNS (OFF, ON, Auto).

•  OFF:  The DNS is disabled.

•  ON: 

The DNS is enabled. Set the domain name, and the DNS server’s primary and 

secondary IP addresses and domain suffixes.

•  Auto:  The DNS is enabled. Set the domain suffixes. The domain name and the DNS server 

IP addresses are set automatically. The Auto option is only displayed when DHCP is 

set to ON.

Содержание WT1800E

Страница 1: ...IM WT1801E 02EN 3rd Edition Precision Power Analyzer WT1801E WT1802E WT1803E WT1804E WT1805E WT1806E ...

Страница 2: ...ower Analyzer Communication Interface User s Manual IM WT1801E 17EN The supplied CD contains the PDF file of this manual This manual explains the communication interface features of this instrument and how to use them WT1801E WT1802E WT1803E WT1804E WT1805E WT1806E Precision Power Analyzer IM WT1801E 92Z1 Document for China The EN and Z1 in the manual numbers are the language codes Contact informa...

Страница 3: ...Adobe and Acrobat are either registered trademarks or trademarks of Adobe Systems Incorporated Modbus is a registered trademark of AEG Schneider In this manual the and TM symbols do not accompany their respective registered trademark or trademark names Other company and product names are registered trademarks or trademarks of their respective holders Revisions 1st Edition September 2016 2nd Editio...

Страница 4: ...s that can be taken to prevent such occurrences CAUTION Calls attention to actions or conditions that could cause light injury to the user or damage to the instrument or user s data and precautions that can be taken to prevent such occurrences French AVERTISSEMENT Attire l attention sur des gestes ou des conditions susceptibles de provoquer des blessures graves voire mortelles et sur les précautio...

Страница 5: ...ds to the appropriate setting F Press the soft key to use the cursor keys to configure this setting After you configure the setting the status of the selected setting switches each time you press the soft key G Press the soft key to execute the specified feature H Press the soft key to apply the value assigned to the key How to Display the Setup Menus That Are Written in Purple below the Keys In t...

Страница 6: ... right cursor keys Changes which digit to set When selecting the item to set Up and down cursor keys Moves the cursor between settings How to Enter Values in Setup Dialog Boxes 1 Use the keys to display the appropriate setup dialog box 2 Use the cursor keys to move the cursor to the item that you want to set 3 Press SET The operation varies as indicated below depending on what you are setting A se...

Страница 7: ...osition you want to insert characters into To switch between uppercase and lowercase letters move the cursor to CAPS on the keyboard and then press SET To delete the previous character press the Back Space soft key To delete all the characters press the All Clear soft key 3 Repeat steps 1 and 2 to enter all the characters in the string Select on the keyboard or press the History soft key to displa...

Страница 8: ...t History A List of Previously Entered Character Strings After selecting an item with the cursor keys press SET to confirm it Enter a preset character string Note cannot be entered consecutively File names are not case sensitive Comments are case sensitive The following file names cannot be used due to MS DOS limitations AUX CON PRN NUL CLOCK COM1 to COM9 and LPT1 to LPT9 For details on file name ...

Страница 9: ...Line Filters 1 17 1 14 Setting Frequency Filters 1 18 1 15 Setting the Data Update Interval 1 19 1 16 Setting Averaging 1 20 1 17 Displaying the Menu for Configuring All Elements 1 21 1 18 Displaying the Setup Parameter List 1 22 Chapter 2 Harmonic Measurement Conditions Option 2 1 Setting Harmonic Measurement Conditions 2 1 Chapter 3 Motor Evaluation Conditions Option 3 1 Setting Motor Evaluation...

Страница 10: ...ctions to Display and the Vertical Scales 10 3 Chapter 11 Bar Graph Display Option 11 1 Setting the Display Format 11 1 11 2 Setting the Measurement Function to Display and the Vertical Scale 11 2 Chapter 12 Vector Display Option 12 1 Setting the Display Format 12 1 12 2 Setting the Element and Wiring Unit to Display and the Zoom Factor 12 2 Chapter 13 Split Display 13 1 Configuring the Split Disp...

Страница 11: ...uring TCP IP Settings 20 3 20 3 Accessing this instrument from a PC FTP Server 20 4 20 4 Monitoring the display of this instrument from a PC Web Server 20 5 20 5 Connecting to a Network Drive 20 8 20 6 Using SNTP to Set the Date and Time 20 9 Chapter 21 Other Features 21 1 Viewing System Information Overview 21 1 21 2 Initializing Settings 21 2 21 3 Setting the Message Menu and USB Keyboard Langua...

Страница 12: ...nt numbers are larger than the selected element On models that have six input elements installed up to three wiring units ΣA ΣB and ΣC are automatically set The wiring unit symbols ΣA ΣB and ΣC are attached to the element numbers in order starting with the smallest number Note Because the wiring system with the largest element number is automatically determined according to the settings of the wir...

Страница 13: ... elements for which the following conditions are met The input element type for 5 A or for 50 A is the same The valid measurement range setting see section 1 7 is the same 2 Follow the instructions below to set the voltage range Auto Range Setting Press the voltage range s AUTO key Setting the Fixed Range Press the voltage range s fixed range keys and to set the voltage range Available Voltage Ran...

Страница 14: ...urrent range Auto Range Setting Press the current range s AUTO key Setting the Fixed Range Press the current range s fixed range keys and to set the current range Available Current Range Options 5 A Input Element When the crest factor is set to 3 When the crest factor is set to 6 or 6A 10 mA 20 mA 50 mA 100 mA 200 mA 500 mA 1 A 2 A 5 A 5 mA 10 mA 25 mA 50 mA 100 mA 250 mA 500 mA 1 A 2 5 A 50 A Inp...

Страница 15: ...n set the current range that is used when current is applied directly to this instrument see section 1 2 3 Follow the instructions below to set the external current sensor range Auto Range Setting Press the current range s AUTO key Setting the Fixed Range Press the current range s fixed range keys and to set the external current sensor range Available External Current Sensor Range Options When the...

Страница 16: ...ion ratio of the input element that is indicated by the cursor is copied to all the input elements in that element s wiring unit Cursor use the cursor keys to move it Note When using the dedicated shunt box you can select an external current sensor conversion ratio preset in the menu for configuring all elements see section 1 17 External Current Sensor Range and Conversion Ratio Configuration Exam...

Страница 17: ...Press SHIFT the ELEMENT ALL key for setting ranges to collectively configure all the input elements for which the following conditions are met The input element type for 5 A or for 50 A is the same The valid measurement range setting see section 1 7 is the same 2 Press EXT SENSOR to illuminate the EXT SENSOR key Press EXT SENSOR again to turn the EXT SENSOR key off 3 Press SHIFT the current range ...

Страница 18: ...he CT ratio Set the power coefficient When you want to multiply the external current sensor output by the conversion ratio and read the current of the circuit under measurement directly turn the VT CT scaling feature off If it is turned on the value will be further multiplied by the CT ratio Display the menu for configuring all elements see section 1 17 Setting the VT Ratio VT Scaling Press the VT...

Страница 19: ... it Note When using the dedicated CT you can select a CT ratio preset in the menu for configuring all elements see section 1 17 Setting the Power Coefficient SF Scaling Press the SF Scaling soft key to display the following menu Set the power coefficient Copies the power coefficient The power coefficient of the input element that is indicated by the cursor is copied to all the input elements in th...

Страница 20: ...ccurs When Element Independent is set to OFF the input elements that are assigned to the same wiring unit are set to the same range If auto range is on you can turn it on by pressing AUTO this instrument operates as follows When a peak over range occurs the measurement range increases to the range specified here skipping the ranges in between When the measurement range to switch to when a peak ove...

Страница 21: ... is displayed in yellow For each input element or wiring unit you can set all ranges as valid measurement ranges All ON Available options for the measurement ranges that this instrument can switch to when a peak over range occurs For each range you can set whether the range is a valid measurement range for all input elements All ON or not All OFF Slot in which a 50 A input element is not installed...

Страница 22: ...measurement ranges whose check boxes have been selected Available current range options Setting the Valid Measurement Range of External Current Sensors Ext Sensor Input Element Valid measurement range The measurement range switches in order between the ranges whose check boxes are selected Ranges whose check boxes are not selected are skipped When Element Independent see section 1 9 is set to OFF ...

Страница 23: ...ystems Set the denominator and numerator of the efficiency equation to the active power and motor power measurement functions P1 to P6 1 PΣA to PΣC 2 Pm 3 Udef1 Udef2 You can set up to four equations η1 to η4 Installed input elements To add active powers and motor output and use them in η1 to η4 use Udef1 and Udef2 Define Udef1 and Udef2 P1 to P6 1 PΣA to PΣC 2 Pm3 1 Can be set within the range of...

Страница 24: ...Off This section explains how to turn the independent input element configuration on and off Independent Input Element Configuration Element Independent in the features guide Wiring Menu Press WIRING to display the following menu Turns independent input element configuration on and off ...

Страница 25: ...omputation Settings Δ Measure Press WIRING and then the Δ Measure soft key to display the following screen The set wiring systems Set the delta computation type The available options vary depending on the set wiring systems Installed input elements Set the delta computation mode rms mean dc r mean ac Wiring System Delta Computation Type 1P3W Difference 3P3W 3V3A 3P3W Difference 3P3W 3V3A 3P4W Star...

Страница 26: ...est Factor This section explains how to set the crest factor Crest Factor Crest Factor in the features guide System Config Menu Press UTILITY and then the System Config soft key to display the following menu Set the crest factor CF3 CF6 CF6A ...

Страница 27: ...e interval is Auto Setting the Synchronization Source for When the Data Update Interval is Auto Sync Source Setting Press SYNC SOURCE and then the Sync Src Setting soft key to display the following screen Set the synchronization source level for voltage current and external current sensor signals When the rectifier function is off 100 0 to 100 0 When the rectifier function is on 0 0 to 100 0 absol...

Страница 28: ...ilter Menu Press LINE FILTER to display the following menu Configure the line filter settings Turn the line filter on or off Set the cutoff frequency 0 1kHz to 100 0kHz in steps of 0 1 kHz 300kHz 1MHz Copies the line filter setting The line filter setting of the input element that is indicated by the cursor is copied to all the input elements in that element s wiring unit Cursor use the cursor key...

Страница 29: ...menu for when the data update interval is Auto Freq Filter A Menu When the data update interval is Auto Press SHIFT LINE FILTER FREQ FILTER and then the Freq Filter at Update Rate Auto soft key to display the following menu Set the frequency filter when the data update interval is Auto Turn the frequency filter on or off Set the cutoff frequency 100 Hz 200 Hz 400 Hz 800 Hz 1 6 kHzv 3 2 kHz 6 4 kHz...

Страница 30: ...menu Increases the data update interval1 Set the data update interval 50ms 100ms 200ms 500ms 1s 2s 5s 10s 20s Decreases the data update interval1 The current data update interval Set whether to set the data update interval to Auto ON OFF Set the timeout2 value for when the data update interval is Auto 1s 5s 10s 20s 1 You can set this when the data update interval is not Auto 2 You can set this whe...

Страница 31: ...e Attenuation constant Average count Averaging AVG in the features guide Averaging Menu Press AVG to display the following menu Set the averaging type Exp Lin Turns averaging on and off Set the attenuation constant or average count When Type is set to Exp Set the attenuation constant 2 to 64 When Type is set to Lin Set the average count 8 to 64 ...

Страница 32: ... Elements All Elements Setup in the features guide All Elements Setup Menu 1 Press WIRING and then the All Elements Setup soft key to display the following menu Use the cursor keys to select the setting that you want to change and then press SET to display the available options or an input box Use the cursor keys to select the item that you want to set ...

Страница 33: ...The INPUT INFO key illuminates and the split display appears The top half of the screen displays the setup parameter list Press INPUT INFO again to clear the setup parameter list and display the previous screen 2 Hold down FORM until the Info Form menu appears Input element or measurement range settings are displayed Select Power Element Settings Input Element Settings List Select Range Settings M...

Страница 34: ...1 23 IM WT18O1E 02EN Info Items Menu 3 Press ITEM to display the Info Items menu Turns the display frame on and off 1 17 Displaying the Setup Parameter List ...

Страница 35: ... harmonic order 0 1 Set the maximum value of the measured harmonic order 1 to 500 Set the distortion factor equation 1 Total 1 Fundamental Set the PLL source U1 I1 U2 I2 U3 I3 U4 I4 U5 I5 U6 I6 Ext Clk Group Hrm2 Set the minimum value of the measured harmonic order 0 1 Set the maximum value of the measured harmonic order 1 to 500 Set the distortion factor equation 1 Total 1 Fundamental Menu on a M...

Страница 36: ...Element Settings Press the Element Settings soft key to display the following menu Set the group of the input element Hrm1 Hrm2 Input elements that are assigned to the same wiring unit are set to the same group 2 1 Setting Harmonic Measurement Conditions ...

Страница 37: ...U1 I1 U2 I2 U3 I3 U4 I4 U5 I5 U6 I6 Ext Clk None Even if Sense Type is set to Pulse correctly setting the synchronization source improves measurement accuracy When Sense Type is set to Pulse Set the upper and lower limits Revolution signal 0 0000 to 99999 9999 rpm Torque signal 10000 0000 to 10000 0000 N m Set the positive and negative rated torque signal pulse frequencies 1 to 100000000 Hz Set th...

Страница 38: ... second X axis value V and Y axis value Nm 1 000 T to 1 000 T Computes A and B Cancels the computation Setting the Electrical Angle Correction Value Electrical Angle Correction On the motor evaluation conditions setup screen select Electrical Angle Correction to display the following screen Set the correction value 180 00 to 180 00 Clears the correction value Automatically computes the correction ...

Страница 39: ... the scaling factor 0 0001 to 99999 9999 Set the unit up to 8 characters Turns the auto range on and off Set the fixed range 20V 10V 5V 2V 1V 500mV 200mV 100mV 50mV Set the linear scale A 1 000 m to 1 000 M B 1 000 M to 1 000M Set A the slope and B the offset Computes A and B Set the line filter OFF 100Hz 1kHz You can configure up to two input signals Computing A and B Calculation Compute A the sl...

Страница 40: ...numeric data that is being printed on the built in printer and communication output are also held Press HOLD again to turn the HOLD key off This releases the hold feature The measured data is then updated at the specified data update rate see section 1 15 Data update count Data update interval If you hold the measured value the data update count in the lower left of the screen stops increasing Cha...

Страница 41: ... performed at the specified data update rate and this instrument then holds the measured value Note If while the HOLD key is illuminated you press HOLD again the HOLD key will turn off and the hold feature will be released If you press SINGLE while the hold feature is released the measured value is updated re measured when the time specified by the data update rate elapses after you press the key ...

Страница 42: ...Info Form menu may be displayed If this happens press FORM again Select the 4 Items display Select the All Items display Select the harmonics list display G5 or G6 option This instrument switches between the single and dual list displays each time you press this soft key Select the custom display You can load the background and customize the numeric data display see section 6 7 Select the 8 Items ...

Страница 43: ...yed page separately for the 4 Items 8 Items 16 Items Matrix All Items and Custom displays For the All Items display the first page is always displayed in the top half of the screen and the currently selected page from pages 2 to 12 is displayed in the bottom half of the screen On the split display you can switch between pages 1 to 12 For the Custom display you can switch between pages when the dis...

Страница 44: ...ay the next page Press SHIFT PAGE to jump to the first page Press SHIFT PAGE to jump to the last page The arrows that can be used to switch the page are displayed in black To the previous page To the next page PAGE PAGE PAGE PAGE Currently displayed page Measurement function side Harmonic order data side To the previous page To the next page PAGE PAGE PAGE PAGE Note If you do not perform step 2 to...

Страница 45: ...list is being displayed the INPUT INFO key is illuminated the Info Items menu may be displayed If this happens press ITEM again In step 1 you can also display the Numeric 4 Numeric 8 or Numeric 16 menu by pressing NUMERIC ITEM and then repeatedly pressing NUMERIC Example of the Numeric 4 Menu Select the item number that you want to set Set the resetting of displayed items Turns the display frame o...

Страница 46: ... or the SHIFT PAGE and keys to select the item that you want to change 5 Press the function select key that corresponds to the measurement function that you want to display Function select keys U I P key S Q λ Φ key WP q TIME key FU FI η key and U I MODE key 6 Press the ELEMENT key for displaying numeric data to select the element and wiring unit that you want to display Press SHIFT the ELEMENT AL...

Страница 47: ...p parameter list is being displayed the INPUT INFO key is illuminated the Info Items menu may be displayed If this happens press ITEM again In step 1 you can also display the Matrix Items menu by pressing NUMERIC ITEM and then repeatedly pressing NUMERIC Select the item number that you want to set 1 to 81 Set the resetting of displayed items Configure the columns to display Turns the display frame...

Страница 48: ...n Element Origin Function Origin Clear Current Page Clear All Pages Column Settings Menu Press the Column Settings soft key to display the following menu Resets items to the default values Set the number of columns 4 6 Set the column number 1 to 6 Set the element and wiring unit None Element 1 to Element 6 ΣA to ΣC 6 4 Changing the Displayed Items on the Matrix Display ...

Страница 49: ... SHIFT PAGE and keys to select the row that you want to change 5 Press the function select key that corresponds to the measurement function that you want to display Function select keys U I P key S Q λ Φ key WP q TIME key FU FI η key and U I MODE key Changing the Element and Wiring Unit Horizontal direction 4 Use the cursor keys to select the column that you want to change 5 Press the ELEMENT key ...

Страница 50: ...c All menu by pressing NUMERIC ITEM and then repeatedly pressing NUMERIC Turns the display of numeric data of all elements or all wiring units on and off If the total number of elements or wiring units is 7 or more set this to ON when you want to display the numeric data of all elements or all wiring units Turns the display frame on and off Set the harmonic order Total 0 to 500 G5 or G6 option You...

Страница 51: ...be displayed If this happens press ITEM again In step 1 you can also display the List Items menu by pressing NUMERIC ITEM and then repeatedly pressing NUMERIC There is a List Items menu for the single harmonics list and the dual harmonics list When you repeatedly press NUMERIC the menu for the single harmonics list is displayed after the All Items display and the menu for the dual harmonics list i...

Страница 52: ...t to display Function select keys U I P key and S Q λ Φ key The WP q TIME key FU FI η key and U I MODE key are disabled 6 Press the ELEMENT key for displaying numeric data to select the element and wiring unit that you want to display If you are displaying the dual harmonics list press SHIFT the ELEMENT ALL key for displaying numeric data to illuminate the indicator below the ELEMENT key and chang...

Страница 53: ...o Custom Custom Items Menu 2 Press ITEM to display the following menu If the setup parameter list is being displayed the INPUT INFO key is illuminated the Info Items menu may be displayed If this happens press ITEM again Loads a display configuration file Turns the display frame on and off Loads a background file Loads display configuration and background files at the same time page 6 14 Configure...

Страница 54: ...ction None other functions for details on the various measurement functions see Items That This Instrument Can Measure in the features guide Set the element and wiring unit Element 1 to Element 6 ΣA to ΣC When Function is set to None Set the character string up to 15 characters When the measurement function includes a harmonic order Set the harmonic order Total 0 to 500 G5 or G6 option Set the dis...

Страница 55: ...me Load Items Bmp Follow the procedure on page 6 12 to display the Custom Items menu and then press the Load Items Bmp soft key to display the following screen Loads the file Select the file that you want to load Select a custom display file Display configuration files txt files Background files bmp files To load both a display configuration file and background file at the same time load the displ...

Страница 56: ... Custom Display Note After you properly load a display configuration file and a background file if you restart this instrument and the same background file is not in the same location the background will return to its default 6 7 Setting the Custom Display ...

Страница 57: ...ed Functions User Defined Function Press MEASURE and then the User Defined Function soft key to display the following screen Displays the setup screen for user defined functions F1 to F5 Displays the setup screen for user defined functions F6 to F10 Displays the setup screen for user defined functions F11 to F15 Displays the setup screen for user defined functions F16 to F20 Turns max hold on and ...

Страница 58: ...rform judgment Range Set the measurement function for details on the various measurement functions see Items That This Instrument Can Measure in the features guide Set the element and wiring unit Element 1 to Element 6 ΣA to ΣC Set the harmonic order Total 0 to 500 G5 or G6 option You can set this setting when the measurement function includes a harmonic order Set the comparison condition OFF Set ...

Страница 59: ... Equations Formula in the features guide Formula Menu Press MEASURE and then the Formula soft key to display the following menu Set the apparent power equation Urms Irms Umean Imean Udc Idc Umean Irms Urmean Irmean Set the apparent power and reactive power equation types Type 1 Type 2 Type 3 Set the corrected power equation Setting the Corrected Power Equation Pc Formula Press the Pc Formula soft ...

Страница 60: ...ing Frequency This section explains how to set the sampling frequency Sampling Frequency Sampling Frequency in the features guide Measure Menu Press MEASURE to display the following menu Set the sampling frequency Auto Clock A Clock B Clock C ...

Страница 61: ...ay Format This section explains how to set the phase difference display format Phase Difference Display Format Phase in the features guide Measure Menu Press MEASURE to display the following menu Set the phase difference display format 180 degrees 360 degrees ...

Страница 62: ...xplains the following setting for master and slave synchronized measurement Master and slave Master Slave Synchronized Measurement Sync Measure in the features guide Measure Menu Press MEASURE to display the following menu Select whether this is the master unit or a slave unit Master Slave ...

Страница 63: ...ng the Voltages or Currents Whose Frequencies Will Be Measured This instrument can measure the frequencies of the voltages or currents of all elements so the setup menu is not displayed even if you press SHIFT MEASURE FREQ MEASURE ...

Страница 64: ... key is displayed when Independent Control is set to ON The input elements that independent integration will be performed on are displayed Turns independent integration on and off Even if you select input elements that independent integration will be performed on independent integration may not be performed due to the wiring system setting or the independent input element configuration see the fea...

Страница 65: ...Continuous Turns integration auto calibration on and off1 Set the watt hour integration methods for each polarity 1 Set the current modes for current integration Set the rated time of integrated D A output DA option Set the integration mode Normal Continuous R Normal R Continuous 1 1 You can set this when the data update interval is not Auto Setting the Integration Timer Integ Timer Press the Inte...

Страница 66: ...dule setup method Each All When you select Each you can set the schedule for each input element Scheduled integration start time Scheduled integration stop time Sets the scheduled integration start time to the current time Copies the scheduled integration start time to the scheduled integration stop time Setting the Watt Hour Integration Method for Each Polarity WP Type Press the WP Type soft key ...

Страница 67: ...de setup method Each All When you select Each you can set the current mode for each input element Setting the Rated Time of Integrated D A Output D A Output Rated Time DA option Press the D A Output Rated Time soft key to display the following screen Set the rated time of integrated D A output 00000 hours 00 minutes 00 seconds to 10000 hours 00 minutes 00 seconds 8 2 Setting Integration Conditions...

Страница 68: ...force integration to stop press this soft key The integration time and integrated value are held The STOP indicator to the right of the INTEG key blinks Integration has stopped Integ Stop is displayed If you press the Start soft key when Stop is displayed in yellow you can resume integration from the point where you stopped integration The STOP indicator to the right of the INTEG key illuminates I...

Страница 69: ...n at Power Failure Recovery Integration Resume Action at Power Failure Recovery Integration Resume Action in the features guide Integration Resume Action Menu Press UTILITY the System Config soft key and then the Preference soft key to display the following menu Set the Integration Resume Action at Power Failure Recovery Start Stop Error ...

Страница 70: ...ivisions of the waveform screen Single Dual Triad Quad Hexa 1 Set waveform mapping Set the time axis 0 05 ms to 2 Configure trigger settings 3 Configure the advanced waveform display settings the specified data update interval 10 1 In addition to using this Format soft key you can repeatedly press WAVE to change the order and the number of divisions 2 For information on how to set the data update ...

Страница 71: ...isplay Settings menu are also reflected in the advanced trend display settings see section 10 1 Setting Waveform Mapping Wave Mapping Press the Wave Mapping soft key to display the following screen Set the mapping destination the divided screen number 0 to 5 Map each waveform U1 I1 etc to the part of the divided screen that you want it to appear on These settings are displayed when Mode is set to ...

Страница 72: ...llowing menu If the setup parameter list is being displayed the INPUT INFO key is illuminated the Info Form menu may be displayed If this happens press ITEM again Set the vertical position 0 000 to 130 000 Set the vertical zoom factor 0 1 0 2 0 25 0 4 0 5 0 75 0 8 1 1 14 1 25 1 33 1 41 1 5 1 6 1 77 2 2 28 2 66 2 83 3 2 3 54 4 5 8 10 12 5 16 20 25 40 50 100 Select the waveforms that you want to dis...

Страница 73: ... you can still show the trend display You can split the screen into top and bottom halves and show the trend display in one half and another display in the other half chapter 13 Trend Form Menu Press FORM to display the following menu If the setup parameter list is being displayed the INPUT INFO key is illuminated the Info Form menu may be displayed If this happens press FORM again Set the number ...

Страница 74: ...owing menu Turns the display of scale values on and off Set the display interpolation Set the grid Turns the display of trend labels on and off Note Changes that you make to the trend display settings on the Display Settings menu are also reflected in the advanced waveform display settings see section 9 1 10 1 Setting the Display Format ...

Страница 75: ...st is being displayed the INPUT INFO key is illuminated the Info Form menu may be displayed If this happens press ITEM again Set the element and wiring unit Element 1 to Element 6 ΣA to ΣC Set the measurement function for details on the various measurement functions see Items That This Instrument Can Measure in the features guide Select the trends that you want to display If you move the cursor to...

Страница 76: ...ll show the bar graph display You can split the screen into top and bottom halves and show the bar graph display in one half and another display in the other half chapter 13 Bar Form Menu Press FORM to display the following menu If the setup parameter list is being displayed the INPUT INFO key is illuminated the Info Form menu may be displayed If this happens press FORM again Set the number of div...

Страница 77: ...EM to display the following menu If the setup parameter list is being displayed the INPUT INFO key is illuminated the Info Form menu may be displayed If this happens press ITEM again Select the bar graph number that you want to set 1 2 3 Set the measurement function U I P S Q λ Φ ΦU ΦI Z Rs Xs Rp Xp Set the element Element 1 to Element6 Select the vertical scale s setup method Fixed Manual Set the...

Страница 78: ...ERS to display the following menu Select Vector Even if you select one of these settings you can still show the vector display You can split the screen into top and bottom halves and show the vector display in one half and another display in the other half chapter 13 Vector Form Menu Press FORM to display the following menu If the setup parameter list is being displayed the INPUT INFO key is illum...

Страница 79: ...Info Form menu may be displayed If this happens press ITEM again If setup parameter list is being displayed the vector that you have set to vector number 1 is displayed in the bottom half of the screen Select the vector number that you want to set 1 2 Set the element and wiring unit Element 1 to Element 6 ΣA to ΣC Set the zoom factor 0 100 to 100 000 Set the zoom factor of fundamental wave U 1 or ...

Страница 80: ...he bottom half of the screen Numeric Wave Bar Vector This feature is available on models with the G5 or G6 option The split screen that you configure last is displayed Form Menu Press FORM to switch between the Form menus of the two screens that you set on the Others menu Configure the settings on each menu Display For Instructions on How to Use the Form Menu See Numeric Sections 6 1 and 6 2 Wave ...

Страница 81: ...Vector Section 12 2 If the setup parameter list is being displayed the INPUT INFO key is illuminated the setup parameter list is displayed in the top half of the screen and the screen that you set to display in the top half of the split screen on the Others menu is displayed in the bottom half of the screen Additionally if you repeatedly press ITEM you can switch between the Info Form menu and the...

Страница 82: ...forms Wave Cursor Menu 2 Press SHIFT FORM CURSOR to display the following menu Set the cursor movement path Max Min Mid Turns the cursor display on and off Set the waveform to measure with cursor 1 U1 I1 U2 I2 U3 I3 U4 I4 U5 I5 U6 I6 Speed 1 Torque 1 Aux1 2 Aux22 Set the waveform to measure with cursor 2 U1 I1 U2 I2 U3 I3 U4 I4 U5 I5 U6 I6 Speed 1 Torque 1 Aux1 2 Aux22 Set the positions of cursor ...

Страница 83: ...nt on and off Cursor Measurement in the features guide 1 Follow the procedures in chapter 10 to display trends Trend Cursor Menu 2 Press SHIFT FORM CURSOR to display the following menu Turns the cursor display on and off Set the trend to measure with cursor 1 T1 to T16 Set the trend to measure with cursor 2 T1 to T16 Set the positions of cursor 1 and cursor 2 0 which is the left edge of the screen...

Страница 84: ...d off Cursor position Turning linked cursor movement on and off Cursor Measurement in the features guide 1 Follow the procedures in chapter 11 to display bar graphs Bar Cursor Menu 2 Press SHIFT FORM CURSOR to display the following menu Turns the cursor display on and off Set the harmonic order of the bar graph to measure with cursor 1 and cursor 2 0 to 500 Turns linked cursor movement on and off ...

Страница 85: ...ess OTHERS to display the following menu Select High Speed Data Capturing HS Settings Menu Press FORM to display the following menu Set the number of data captures Infinite 1 to 9999 Confirm and optimize the maximum capturing count Starts high speed data capturing section 15 4 Stops high speed data capturing section 15 4 Set the save conditions section 15 2 Select whether to save to a file OFF Whe...

Страница 86: ...to the maximum capturing count displayed to the left Configuring Capture Control Settings Control Settings Press the Control Settings soft key to display the following menu Turn external signal synchronization on or off OFF This instrument captures data every 5 ms ON This instrument captures data in sync with an external signal Configure trigger settings Configure the voltage and current measureme...

Страница 87: ...gs Set the trigger slope Set the trigger mode Auto Normal OFF Set the trigger source U1 I1 U2 I2 U3 I3 U4 I4 U5 I5 U6 I6 Ext Clk Set the trigger level 0 0 to 100 0 15 1 Setting the Number of Data Captures and Configuring the Capture Control Settings ...

Страница 88: ...Comment Save Conditions File Settings in the features guide 1 Follow the procedure in section 15 1 to select High Speed Data Capturing on the Others menu File Settings Menu 2 Press FORM and then the File Settings soft key to display the following menu Set the save destination and perform manual CSV conversion Set auto naming section 17 2 Select whether to perform automatic CSV conversion when high...

Страница 89: ... ON or not All OFF Slot in which an input element is not installed You can set whether to save all the numeric data All ON or not All OFF For each input element or wiring unit you can set whether to save the numeric data for U I and P All ON or not All OFF You cannot configure this setting for slots that do not have input elements installed in them On models with the MTR option you can set whether...

Страница 90: ...ection 15 1 to select High Speed Data Capturing on the Others menu HS Items Menu 2 Press ITEM to display the following menu Resets items to the default values Peak over range information Turns the display frame on and off Set the number of columns 4 6 Set the column number 1 to 6 Set the element or wiring unit None Element 1 to Element 6 ΣA to ΣC Switching the Page You can switch between page 1 an...

Страница 91: ...n models with the MTR option Aux1 and Aux2 are displayed on models with the AUX option Page 4 Example Page 4 can only be selected on models with the MTR or AUX option Speed Torque and Pm are displayed on models with the MTR option Aux1 and Aux2 are displayed on models with the AUX option 15 3 Changing the Displayed Items for High Speed Data Capturing ...

Страница 92: ...elect the column that you want to change 5 Press the ELEMENT key for displaying numeric data to select the element and wiring unit that you want to display In high speed data capturing the elements and wiring unit configurations are the same on all four pages If you change the element and wiring unit configuration on one page the configuration changes on the other pages as well ELEMENT key for dis...

Страница 93: ...turing HS State Start appears in the upper right of the screen While the captured data is being saved File State Rec appears in the upper left of the screen French ATTENTION Pendant la collecte de données haute vitesse et lorsque les données collectées sont enregistrées le système a constamment accès au support de stockage même si l icône qui l indique n est pas affichée Ne retirez pas le support ...

Страница 94: ...creen When the captured data is being saved File State Rec appears in the upper left of the screen Stops high speed data capturing After the specified number of data captures have been made high speed data capturing automatically stops To force high speed data capturing to stop press this soft key When high speed data is stopped HS State Ready appears in the upper right of the screen Note You cann...

Страница 95: ...Optimizing the Maximum Storage Count Press the Optimize Count soft key to display the following screen Maximum storage count 0 to the maximum number of times that data can be stored to the save destination The maximum number of times that storage can be performed depends on the number of stored items that you have set and the free space at the save destination For details on how to set the stored ...

Страница 96: ...the Real time Control soft key to display the following screen Set the scheduled start and stop times Year month day 00 hours 00 minutes 00 seconds to 23 hours 59 minutes 59 seconds Scheduled storage start time Scheduled storage stop time Sets the scheduled storage start time to the current time Copies the scheduled storage start time to the scheduled storage stop time Integration Synchronized Sto...

Страница 97: ...ge count page 16 1 Select the trigger event Event 1 to Event 8 When measured data is updated storage is started if the conditions of the specified user defined event are met Single Shot Storage Mode Set Store Mode to Single Shot Set the storage count Infinite 1 to 9999999 Confirms and optimizes the maximum storage count page16 1 16 1 Configuring Storage Control ...

Страница 98: ...ayed on the screen Stores the numeric data that you have specified on the Store Items setup screen Press this soft key to store the numeric data items that you have specified on the Store Items setup screen that is displayed when you press the Items soft key Set the stored items This soft key is displayed when you select Selected Items Setting Stored Items Items Press the Items soft key to display...

Страница 99: ...d perform manual CSV conversion Set auto naming section 17 2 Select whether to perform automatic CSV conversion when storage stops Set the file name section 17 2 Set a comment section 17 2 Setting the Save Destination and Performing Manual CSV Conversion Press the File List soft key to display the following screen Set the save destination When storage starts see section 16 4 storage data is saved ...

Страница 100: ...ckage est en cours quand la touche STORE START est éclairée ou quand elle clignote ou bien quand la touche STORE STOP clignote Starting the Storage Operation Press STORE START This instrument starts storage using the storage mode that you have specified see section 16 1 The STORE START key is illuminated Storage has started Store Start is displayed The STORE START key is blinking The storage opera...

Страница 101: ...g stored data to a file and closes the file If the storage operation has automatically stopped When the storage operation stops automatically this instrument finishes writing stored data to a file and closes the file Therefore the reset operation performs no file operations Note If you do not reset storage you will not be able to start it again 16 4 Starting Stopping and Resetting Storage ...

Страница 102: ... d endommager le support de stockage ou les données qu il contient Quand le système accède au support de stockage USB s affiche au centre dans la partie supérieure de l écran et le voyant du support de stockage USB clignote USB Memory Devices That Can Be Used and How to Connect USB Memory Devices Use portable USB memory devices that are compatible with USB Mass Storage Class version 1 1 Connect US...

Страница 103: ...ass version 1 1 Do not connect an incompatible USB memory device You cannot use protected USB memory devices such as those that contain encrypted content Do not connect and disconnect the two USB devices repetitively Provide at least a 10 second interval between removal and connection General USB Handling Precautions Follow the general handling precautions that are included with your USB memory 17...

Страница 104: ...te Set a comment Set the file name You can set the file name that is used when Auto Naming is set to OFF This is also used as the common file name when Auto Naming is set to Numbering Saves the data Setting the Save Destination File List Press the File List soft key to display the following screen Operation menu The destination storage medium drive or folder Use the cursor keys to move the cursor ...

Страница 105: ...the exact same date and time down to seconds exists The sequence number that comes after the date and time is appended if a file name with the exact same date and time down to seconds exists The sequence number is incremented by one 0 to 9 and then A to Z each time a file is added Setting the File Name File Name You can set the file name that is used when Auto Naming is set to OFF This is also use...

Страница 106: ...Save destination Auto naming File name Comment Saving Waveform Display Data Save Wave in the features guide Save Wave Menu Press FILE and then the Save Wave soft key to display the following menu Set the save destination section 17 2 Set auto naming section 17 2 Set a comment section 17 2 Set the file name section 17 2 Saves the data ...

Страница 107: ...et a comment section 17 2 Set the file name section 17 2 Saves the data Setting the Numeric Data Items to Save Item Settings Press the Item Settings soft key to display the following menu Saves the numeric data that is displayed on the screen Press this soft key to save the numeric data items that are displayed on the screen Saves the numeric data that you have specified on the File Items setup sc...

Страница 108: ...ed Items soft key on the menu on the previous page the numeric data items that you have specified on the following screen are saved Selects all the numeric data items Clears the selection of all the numeric data items Selects the preset numeric data items Select the check boxes for the numeric items to save 17 4 Saving Numeric Data ...

Страница 109: ...d then the Load Setup soft key to display the following screen Select the SET file that you want to load Loads the file Note This instrument cannot load setup parameters that have been saved by a product with an incompatible firmware version This instrument cannot load setup parameters that were saved by an instrument with a different element configuration or with different options ...

Страница 110: ...eys to move the cursor Sorts the file list Set the display format Set the type of file to list Changes the storage medium drive Deletes the selected files and folders Renames files and folders Makes folders directories Copies the selected files and folders Moves the selected files and folders Total number of files and folders that are contained within the storage medium or folder indicated by the ...

Страница 111: ...name in descending order Sorts by file size in ascending order Sorts by file size in descending order Sorts by date and time in ascending order Sorts by date and time in descending order Setting the Display Format Select or on the operation menu to display the following screen List Display Set the display format Thumbnail Display Set the display format 17 6 File Operations ...

Страница 112: ...isplay background files PNG Screen image data files PNG format JPG Screen image data files JPEG format TXT Custom display configuration files Changing the Storage Medium or Drive Change Drive Select Change Drive on the operation menu to display the following screen Changes the storage medium drive Storage Medium Drive Type RAM 0 Internal RAM disk of this instrument USB 0 USB memory device that was...

Страница 113: ...ut the following operations Files Press SET or the Set Reset soft key on the Utility menu Folders Press the Set Reset soft key on the Utility menu If you press SET all the files and folders that you have selected up to that point will be cleared You cannot delete folders on network drives Renaming Files and Folders Rename 1 Select the file or folder in the file list that you want to rename 2 Selec...

Страница 114: ...es and folders Copy Executes the copy operation on the selected files and folders Copy Exec Folder that you are copying to File list that you are copying to File list that you are copying from 3 Select the drive or folder in the file list that you are copying to 4 Select Copy Exec on the operation menu to display the following screen Confirms the copying of files and folders Note The procedure for...

Страница 115: ...you are moving data from 3 Select the drive or folder in the file list that you are moving to 4 Select Move Exec on the operation menu to display the following screen Confirms the moving of files and folders Note The procedure for selecting multiple files or folders at the same time to move them is the same as the procedure for selecting multiple files or folders at the same time to delete them Fo...

Страница 116: ... key selects the file or folder in the file list that is highlighted or clears the selection The selection marks see page 17 8 are displayed to the left of the selected files Select All and Clear All All Set and All Reset All Set In the file list when a drive is highlighted or a file or folder in a drive or folder is highlighted pressing this soft key selects all the files and folders in the corre...

Страница 117: ...Images in the features guide Image Save Menu Press SHIFT IMAGE SAVE MENU to display the following menu Set the save destination section 17 2 Set the data format BMP PNG JPEG Set the color OFF Color Reverse Gray Set auto naming section 17 2 Set a comment section 17 2 Set the file name section 17 2 Saving Screen Captures Press IMAGE SAVE to save the screen image with the save conditions that you spe...

Страница 118: ...d dark place Use the paper as quickly as possible after you break its protective seal If you attach a plastic film that contains plasticizing material such as vinyl chloride film or cellophane tape to the paper for a long time the recorded sections will fade due to the effect of the plasticizing material Use a holder made of polypropylene to store the roll paper When pasting the record paper to an...

Страница 119: ... dans l ouverture de sortie du papier en rouleau de l imprimante Ne pas laisser une main ou des doigts entrer en contact avec le dispositif de coupe lors de l ouverture du couvercle de l imprimante et de la charge du papier en rouleau dans son logement Ne touchez pas la tête d impression ni le moteur d impression Vous pourriez vous brûler 1 Slide the lever to the right to make the printer unit pro...

Страница 120: ...ment until you hear a click Close the cover until you hear a click Tray Cover Stopper Push the stopper to the left to release the latch Cover Push the printer unit until you hear a click This completes the procedure for loading the roll paper Lever 5 Pull approximately 10 cm of the roll paper out and load the roll paper in the holder so that the thermal side of the paper is facing up Load the pape...

Страница 121: ...the paper up against the top of the cover Note If you open the printer cover immediately after you cut the roll paper repeat steps 5 to 7 on pages 19 2 and 19 3 After you load roll paper and close the cover check whether the paper feeds correctly If the roll paper does not feed straight repeat steps 1 to 7 on pages 19 2 and 19 3 If you load the roll paper backwards the paper may not feed properly ...

Страница 122: ...val scheduled times for real time printing trigger event synchronized to a user defined event and printing data when printing starts Comment Feeding paper Printing Screen Images and Numeric Data Option in the features guide Print Menu Menu Press SHIFT PRINT MENU to display the following menu Feeds paper section 19 1 Set a comment section 17 2 Configure auto printing Turns auto printing on executes...

Страница 123: ...ount Infinite 1 to 9999 Set the print interval Select whether to print the data at print start Setting the Print Interval Press the Print Interval soft key to display the following screen Set the print interval 00 hours 00 minutes 10 seconds to 99 hours 59 minutes 59 seconds Scheduled Times for Real Time Print Mode Set Print Mode to Real Time Set the print count Infinite 1 to 9999 Set the print in...

Страница 124: ...me Sets the scheduled print start time to the current time Copies the scheduled print start time to the scheduled print stop time Integration Synchronized Print Mode Set Print Mode to Integ Sync Set the print interval Interval Print Mode on the previous page Select whether to print the data at print start Event Synchronized Print Mode Set Print Mode to Event Set the print count Infinite 1 to 9999 ...

Страница 125: ...19 8 IM WT18O1E 02EN 19 2 Printing Using the Built in Printer Option Printing Press Print to print data according to the conditions specified in the Print Menu menu ...

Страница 126: ...anical specifications IEEE802 3 Transmission system Ethernet 1000BASE T 100BASE TX 10BASE T Communication protocol TCP IP Supported services FTP server DHCP DNS remote control VXI 11 SNTP FTP client Modbus TCP server and Web server Connector type RJ 45 connector Ethernet port RJ 45 modular jack Items Required to Connect this instrument to a Network Cable Use one of the following types of network c...

Страница 127: ...ument and the PC 2 Connect one end of a UTP or STP cable to the ETHERNET 1000BASE T port on the rear panel 3 Connect the other end of the UTP or STP cable to a hub or router 4 Connect the PC to the hub or router in the same way 5 Turn on this instrument Hub or router that supports 1000BASE T UTP or STP cable straight cable PC WT Note Use a hub or router that supports the data rate of your network ...

Страница 128: ...owing screen Set the DHCP OFF ON Set these when DHCP is set to OFF IP address Subnet mask Default gateway These are displayed when DNS is set to ON or Auto Domain name DNS server IP address 1 primary 2 secondary Domain suffix 1 primary 2 secondary Applies the settings Set the DNS OFF ON Auto OFF The DNS is disabled ON The DNS is enabled Set the domain name and the DNS server s primary and secondar...

Страница 129: ...Web Server Press UTILITY the Network soft key and then the FT WebP Server soft key to display the following menu Set the user name up to 32 characters Set the password up to 32 characters Set the timeout value 30 to 3600 s Applies the settings FTP Client Software Start an FTP client on a PC Enter the user name and password that you entered on the screen shown above to connect to this instrument No...

Страница 130: ...rument from the PC User name Password Connecting to the DLM4000 from a PC Web Server Web Server in the features guide Configuring Web Server Settings FTP Web Server Press UTILITY the Network soft key and then the FTP Web Server soft key to display the following menu Set the user name up to 32 characters Set the password up to 32 characters Applies the settings Note Time Out is a setting used by th...

Страница 131: ...is shown on the previous page and connect to this instrument The following screen appears Note If you set the user name to anonymous you can access this instrument without entering a password Home Screen Information about the instrument is displayed Control Screen 4 Click the Control tab The following screen appears Set the resolution Select LOW 512 384 or HIGH 1024 768 Message Refresh screen butt...

Страница 132: ...ks Screen 5 Click the Link tab The following screen appears Link to Yokogawa Test Measurement power analyzer website Link to Yokogawa Test Measurement website 20 4 Monitoring the display of this instrument from a PC Web Server ...

Страница 133: ...network drive Network Drive Net Drive in the features guide Configuring Network Drive Net Drive Settings and Connecting to It Press UTILITY the Network soft key and then the Net Drive soft key to display the following menu Set the FTP server s host name Set the login name up to 32 characters Set the password up to 32 characters Turns FTP passive mode on and off Set the timeout value 1 to 3600 s Di...

Страница 134: ...he System Config menu Time adjustment SNTP SNTP in the features guide Configuring SNTP Settings SNTP Press UTILITY the Network soft key and then the SNTP soft key to display the following menu Configure the SNTP server settings IP address host name and domain name can be set when DNS is enabled Set the timeout value 1 to 60 s Turns automatic adjustment on and off Set the time difference from Green...

Страница 135: ...mation List System Overview Press UTILITY and then the System Overview soft key to display the following screen Displayed Contents Model Model Suffix Suffix code No Instrument number Version Firmware version Element Configuration Input element types Options Options Link Date Date and time that the firmware was created Product ID Unique number assigned to each instrument necessary for the purchase ...

Страница 136: ...n the features guide Utility Menu Press UTILITY to display the following menu Initializes the settings Confirmation to Initialize Settings Note Only initialize this instrument if you are sure that it is okay for all of the settings to be returned to their initial values You cannot undo an initialization We recommend that you save the setup parameters before you initialize this instrument ...

Страница 137: ... soft key to display the following menu Configure the language settings Set the USB keyboard language Japanese English Set the date and time section 3 4 in the getting started guide IM WT1801E 03EN Configuring the Language Settings Press the Language soft key to display the following menu Set the menu language Japanese English Chinese German Set the message language Japanese English Chinese German...

Страница 138: ...rn off the LCD backlight Set the amount of time before the backlight is automatically turned off 1min to 60min Set the LCD backlight brightness 1 to 10 Configure the display color settings Turns off the LCD backlight you can turn on the backlight by pressing any key Configuring the Display Color Settings Press the Color Settings soft key to display the following menu Sets the graph colors to the d...

Страница 139: ...t Settings Preference in the features guide Preference Menu Press UTILITY the System Config soft key and then the Preference soft key to display the following menu On models with the MTR option set the motor display value when the measured pulse frequency is less than the lower limit 0 Error Set the decimal point and separator to use when data is saved in ASCII format as a CSV file Period Comma Se...

Страница 140: ... to ΣC Set the measurement function None other functions for details on the various measurement functions see Items That This Instrument Can Measure in the features guide Output item This display changes according to the Function Element Σ and Order settings Set the harmonic order Total 0 to 500 G5 or G6 option You can set this setting when the measurement function includes a harmonic order Select...

Страница 141: ...is instrument are functioning properly Test item Memory test Key test Operation keys indicators and keyboard Self Test Selftest in the features guide Selftest Menu Press UTILITY and then the Selftest soft key to display the following menu Set the item to test Memory Key Board Executing the Memory Test Set the item to test to Memory Executes the test ...

Страница 142: ...EN Executing the Key Test Set the item to test to Key Board Displays the keyboard Executes the key test This self test tests the panel keys and indicators Executing the Keyboard Test 21 7 Carrying Out Self Tests Selftest ...

Страница 143: ...nge the measurement range or input filter To make accurate measurements we recommend that you execute zero level compensation after warming up the instrument for at least 30 minutes If the measurement range and input filter remain the same for a long period of time the zero level may change due to the changes in the environment If this happens we recommend that you execute zero level compensation ...

Страница 144: ...ate of the NULL feature for all the signals of a given signal type ON Hold OFF For the voltage signals of the installed input elements For the current signals of the installed input elements For motor evaluation input signals For external signals Set the state of the NULL feature for each signal ON Hold OFF The motor evaluation input signal setup screen is displayed on models with the MTR option T...

Страница 145: ...features guide Key Lock KEY LOCK Press SHIFT LOCAL KEY LOCK LOCK is displayed in the upper right of the screen and the operation keys are locked The key lock disables all keys of this instrument except for the power switch SHIFT key and LOCAL key Press SHIFT LOCAL KEY LOCK again to release the key lock Note When the keys are locked you cannot use a USB mouse or keyboard to operate this instrument ...

Страница 146: ...TOR AUX are switched to fixed ranges Even if the Data Update Interval setting is Auto Voltage Current measurement range are also switched to fixed ranges 1 2 and 1 15 89 Processing system settings change Please wait for a moment 90 This model has no external current sensor Check the specifications to see whether or not the optional external current sensor is provided 21 1 91 This model has no buil...

Страница 147: ...vice is required 621 File is damaged Check the file 643 to 653 Media failure Check the medium 657 File operation is interrupted 658 File unknown format Check the file format 17 5 and 17 6 662 Cannot load this bitmap file Use file of 16bit Color or 24bit Color Mode with less or equal size 800x672 6 7 663 Cannot load this text file Confirm the contents of file 6 7 665 Cannot load this file format Fi...

Страница 148: ...not allowed or the file name is not a valid MS DOS file name Enter another file name 17 2 802 Cannot be set or executed in the Normal measurement mode Usable measurement mode are as follows 811 Cannot be set to this display mode Harmonics option is necessary 812 Cannot be set or executed while storing data 813 Cannot be set while integration is running Reset Integration 8 3 814 Cannot be set or ex...

Страница 149: ...Print Settings menu 19 2 872 Auto print s print out has been canceled The printer or file system is in action 19 2 874 Sync source PLL source or trigger source cannot be set to Ext Clk while Master Slave Synchronization Measurement is set to Slave 7 6 875 Master Slave Synchronization Measurement cannot be set to Slave while sync source PLL source or trigger source is set to Ext Clk 7 6 876 Can not...

Страница 150: ... Maintenance service is required to replace the back up battery 909 Illegal SUM value Maintenance service is required 910 This operation is prohibited for EEPROM protection 915 EEPROM SUM error EEPROM may be damaged Maintenance service is required 919 Module installation condition and setup parameters do not match The system has been initialized Maintenance service is required 920 SUM error of NUL...

Страница 151: ...event synchronized printing 19 7 event synchronized storage 16 3 expression 7 1 7 3 extension 17 11 external current sensor conversion ratio 1 5 external current sensor range 1 4 external VT CT 1 7 EXT SENSOR key 1 4 1 6 EXT SENSOR SENSOR RATIO key 1 5 Index Symbols Page 4 8 and 16 Items displays 6 4 1000BASE T port 20 1 A Page A and B computation auxiliary input 4 1 A and B computation motor eval...

Страница 152: ...nchronized storage 16 2 integration timer 8 2 interval printing 19 6 ITEM key 1 23 6 4 6 6 6 9 6 10 6 12 9 3 10 3 11 2 12 2 13 2 15 6 J Page judgment condition 7 2 K Page k and K iii keyboard operating vi key lock 21 11 keys operating iv key test 21 8 L Page language 21 3 linear scale auxiliary input 4 1 linear scale motor evaluation 3 1 line filter 1 17 line filter auxiliary input 4 1 LINE FILTER...

Страница 153: ...g 1 22 setup parameters loading 17 8 setup parameters saving 17 3 SINGLE CAL key 21 9 SINGLE key 5 2 single measurement 5 2 single shot storage 16 3 SNTP 20 9 sort files 17 10 split display 13 1 S Q λ Φ key 6 5 6 8 6 11 standard corrected power 7 3 storage count 16 1 storage interval 16 1 storage media drive files 17 11 storage mode 16 1 storage numeric data 16 4 storage scheduled times for 16 2 s...

Страница 154: ...veform mapping 9 2 WAVE key 9 1 9 3 Web Server 20 5 WIRING key 1 1 1 12 1 13 1 14 wiring pattern 1 1 wiring system 1 1 wiring unit 1 1 WP q TIME key 6 5 6 8 Z Page zero level compensation 21 9 zoom factor vector 12 2 zoom factor waveform 9 3 Index ...

Отзывы: