IM 253401-01E
8-6
8.4
Setting the Harmonic Order and Displaying the
Results of Harmonic Analysis
Relevant Keys
SCALING
AVG
FILTER
STORE
RECALL
HARMONICS
SAMPLE
V OVER
A OVER
MODE
RMS
V MEAN
DC
A
B
C
hour
hour
min
min
sec
V
VA
m
A
k
var
M W
TIME
V
PF
m
A
k
deg
M W
%
FUNCTION
AUTO
AUTO
MODE
1
Φ
3W
V
Hz
m
A
k
h
M W
TRIG
V RANGE A RANGE
HOLD
ENTER
INTEGRATOR
START
HARMONICS
MEMORY
INTEG SET
STOP
RESET
REMOTE
INTERFACE
OUTPUT
LOCAL
SETUP
h
SHIFT
WIRING
3
Φ
3W
3
Φ
4W
3V3A
ELEMENT
1
2
3
FUNCTION
ELEMENT
1
2
3
FUNCTION
ELEMENT
1
2
3
*Shows the operation panel of the WT130. For the differences
between WT110 and WT130, refer to section 2.2, page 2-2, 2-3
Displays
relevant
keys and
indicator
Operating Procedure
The following operations assume that the harmonic analysis function is turned ON.
Setting the Harmonics Order
1. Light up the display function indicator of display A.
Display
A
Light up display function
V
A
W
FUNCTION
FUNCTION
FUNCTION
FUNCTION
2. Set the harmonics order.
(Display A)
Displaying the Values of Harmonic Analysis
Displays each analysis value after having set the display function of either display B or C.
FUNCTION
V
A
W
PF
FUNCTION
FUNCTION
FUNCTION
FUNCTION
FUNCTION
V% A%
W%
Vdeg
A%
V%
Adeg
FUNCTION
FUNCTION
FUNCTION
FUNCTION
FUNCTION
Display
B
FUNCTION
V
A
W
V Hz
FUNCTION
FUNCTION
FUNCTION
FUNCTION
A Hz
Display
C
Explanation
Setting the Order of Harmonics
The maximum order for which analysis results can be displayed varies depending on the
frequency of the fundamental.
Example
• When the fundamental frequency is 50Hz, up to 50 orders can be displayed;
• When the fundamental frequency is 400Hz, up to 30 orders can be displayed.
When an order is set exceeding the maximum order, display B will change to the dot display.
Refer to Ch. 15 for more details on upper limits of analysis orders.