IM 253401-01E
7-4
7.2
Setting Integration Mode and Integration Timer
Relevant Keys
SCALING
AVG
FILTER
STORE
RECALL
HARMONICS
SAMPLE
V OVER
A OVER
MODE
RMS
V MEAN
DC
A
B
C
hour
hour
min
min
sec
V
VA
m
A
k
var
M W
TIME
V
PF
m
A
k
deg
M W
%
FUNCTION
AUTO
AUTO
MODE
1
φ
3W
V
Hz
m
A
k
h
M W
TRIG
V RANGE A RANGE
HOLD
ENTER
INTEGRATOR
START
HARMONICS
MEMORY
INTEG SET
STOP
RESET
REMOTE
INTERFACE
OUTPUT
LOCAL
SETUP
h
SHIFT
WIRING
3
φ
4W
3
φ
3W
3V3A
ELEMENT
1
2
3
FUNCTION
ELEMENT
1
2
3
FUNCTION
ELEMENT
1
2
3
*Shows the operation panel of the WT130. For the differences
between WT110 and WT130, refer to section 2.2, page 2-2, 2-3
Displays
relevant
keys and
indicator
Operating Procedure
SHIFT
RESET
INTEG SET
ENTER
3.
(Display C)
ENTER
5.
Selecting the mode
(Display C)
2.
4.
End
1.
• Selecting the Integration Mode
• Perform operations following the thick line in the below menu.
• Press the ENTER key to confirm the selection or setting.
• When you want to leave the current menu during operation, press the key described
under step 1. The confirmed settings made until that point will be kept.
hour
min
1.
SHIFT
RESET
INTEG SET
ENTER
3.
(Display C)
Setting the timer
(Display A)
2.
ENTER
6.
4.
Up/down
5.
Cursor shift
SHIFT
End
• Setting the Integration Timer
Explanation
Selecting the Measurement Mode
The following selections are available. The initial value is nor.
• nor :Select this for manual or standard integration mode. Depending on the integration timer,
this instrument will automatically decide the appropriate mode.
• Cont :Select this for the continuous integration mode.
Setting the Integration Timer
This setting decides how long integration will be performed in terms of hours and minutes. The
setting ranges from 000.00 (0 hrs, 0 min) to 999.59 (999 hrs, 59 min). The initial value is
000.00.
• 000.00 :When "nor" is selected on the integration menu, the manual integration mode will
become valid. When "Cont" is selected, an error code will appear and integration will not be
performed.
• 000.01 to 999.59 :The time during which integration will be performed when in the standard
or continuous integration mode. The standard or continuous mode should be selected at the
integration mode menu.