5-1
IM 707281-01E
Specifications
5
Chapter 5 Specifications
5.1
Performance Specifications
The following performance specifications are attained under standard operating
conditions (section 5.3 “General Specifications”).
Common
Number of output channels
4 channels
Output format
Floating unbalanced output, isolated between channels
D/A resolution
16 bits (includes the sign)
Output range
±
1 V,
±
2 V,
±
5 V,
±
10 V
Maximum output current
±
10 mA (per channel)
Allowable load resistance
1 k
Ω
or more
Output impedance
1
Ω
or less
During DC Output
DC accuracy
(at ambient temperature of 23
±
5
°
C, ambient humidity of 50
±
10%RH, after the warm-up
time has passed)
Range Resolution DC Accuracy
Temperature Coefficient (5-18
°
C, 28- 40
°
C)
±
1 V
0.1 mV
±
(0.05% of 0.5 mV)
±
(50 ppm of 33
µ
V)/
°
C
±
2 V
1 mV
±
(0.05% of 1 mV)
±
(50 ppm of 54
µ
V)/
°
C
±
5 V
1 mV
±
(0.05% of 2 mV)
±
(50 ppm of 116
µ
V)/
°
C
±
10 V 1 mV
±
(0.05% of 4 mV)
±
(50 ppm of 220
µ
V)/
°
C
Settling time
32
µ
s (until the value settles within
±
0.1% of the final value, typical vlaue*)
Trigger source
Manual, bus trigger signal (BUSTRG1/BUSTRG2) of the measuring station.
During Function Waveform (FG) Output
Output waveform
Sine, pulse (variable duty cycle), ramp, triangular, arbitrary, and DC.
Output frequency range and resolution
1 mHz to 20 kHz (Sine wave), 1 mHz to 10 kHz (other waves)
Resolution: 1 mHz
Output frequency accuracy
±
(0.01% of s 23
µ
Hz)
maximum output voltage
Range withing the output range
Amplitude range and accuracy
(at ambient temperature of 23
±
5
°
C, ambient humidity of 50
±
10%RH, after the warm-up
time has passed.)
Range Setting
Selectable range
Resolution
Accuracy (for 1-kHz sine wave)
±
1 V
0 to 2 Vp-p
0.1 mVp-p
±
4 mVp-p
±
2 V
0 to 4 Vp-p
1 mVp-p
±
8 mVp-p
±
5 V
0 to 10 Vp-p
1 mVp-p
±
20 mVp-p
±
10 V
0 to 20 Vp-p
1 mVp-p
±
40 mVp-p
Offset voltage range
and accuracy
Range Setting
Selectable Range
Resolution
Accuracy
±
1 V
–1 to 1 V
0.1 mV
±
2 mV
±
2 V
–2 to 2 V
1 mV
±
4 mV
±
5 V
–5 to 5 V
1 mV
±
10 mV
±
10 V
–10 to 10 V
1 mV
±
20 mV