roubleshooting and Maintenance
4
Chapter 4 Troubleshooting and Maintenance
4.1
Troubleshooting
• If servicing is necessary, or if the instrument is not operating correctly after performing
the following corrective actions, contact your nearest YOKOGAWA dealer.
• To verify that the module is operating correctly, perform the self test as described on
the next page.
Problem
Probable Cause/Corrective Action
Reference
Module does not operate.
Check to see that the module is installed correctly into the station.
2-1, 3-1, *
Also, install the module into another slot, and check whether it will
operate there. If it operates in the other slot, the measuring station
is likely to have malfunctioned.
If the module is installed correctly and does not operate, the
connector might be bad or the IC may have malfunctioned. In either
case, contact your nearest YOKOGAWA dealer to have it repaired.
There is no waveform output
Check whether the “O” indicator on the left of the [Output] button
3-15
is green. If the waveform is a trigger oscillation or gate oscillation,
check that the settings are correct.
3-2, 3-4, 3-10
The Output waveform is not correct
Check to see that the output waveform setting and output conditions
3-1 to 3-15
are correct.
Cannot set the waveform
Check that the settings you are trying to enter are within the
5-1 to 5-3
specifications.
* See WE7000 User’s Manual (IM 707001-01E).