<2. Analyzer Preparation>
2-3
IM 12Y18A03-01EN
1st Edition : Oct. 04, 2019-00
n
Sample Requirements
Additional information is listed in the Appendix, Specifications section.
Sample inlet connection
1/4” NPTF. If particulate matter is present in the sample, pre-filtration is necessary. The
60 micron stainless steel filter located after inlet valve will remove moderate amounts of
particulates.
Flow rate
40 mL/minute (nominal).
Pressure
8 to 100 psig.
Temperature
Temperature must be between 5 and 45°C.
Oxygen scavenger level
Oxygen scavenger levels are read directly in ppb or ppm, when calibrated with standard.
Sample alkalinity
Sample alkalinity should be less than 50 ppm CaCO
3
equivalent.
2.2
Electrical Wiring
The warning icon highlights important information that should be strictly followed when using the
analyzer for your own safety. Failure to follow these instructions may result in injuries.
WARNING
Read and observe the following safety recommendations.
n
Safety Requirements
• Prior to wiring, a switch or circuit breaker for disconnecting the analyzer from power supply
should be installed.
• The switch should be in close proximity to the analyzer and with easy reach of the
user.
• The switch should be marked as the disconnecting device for the analyzer.
• To reduce the risk of shock hazard, disconnect the power prior to opening the analyzer.
• Before connecting the analyzer to the main, make sure that the voltage lies within either
range: 100-120V 200mA / 200-240V 100mA; 50/60 Hz AC.
• Cutting off the power by disconnecting power source will not reset the analyzer. This
analyzer incorporates a non-volatile memory and will maintain calibration and settings after
power failure. Battery power is supplied to the display for the date and time functions.
• If a repair is required, or to arrange Return Material Authorization, call Response Center or
contact your local authorized dealer.
• Installation and wiring of the analyzer may only be carried out in accordance with applicable
local and national codes per this user guide.
• Be sure to observe the technical specifications and input ratings.
n
Warning Labels and Locations
WARNING
The following section provides important information that should be strictly followed when using
the analyzer for your own safety. Failure to follow these instructions may result in injuries.