<2. Analyzer Preparation>
2-1
IM 12Y18A03-01EN
1st Edition : Oct. 04, 2019-00
2. Analyzer Preparation
WARNING
The instructions provided in this userguide are recommendations from the manufacturer to
ensure safe and correct operation of the analyzer. If the analyzer is not used as recommended by
the manufacturer this can lead to incorrect operation or injury.
n
Unpacking the Analyzer
YOKOGAWA analyzers are assembled, tested and packaged with great care.
1. Open the outer box. Remove the top four foam corner support pieces.
2. Open the inner box. This box should contain the analyzer and accessories. (Refer to IM
12Y18A00-01EN)
3. Remove the cardboard retaining shell by sliding it over the entire mounting board and the
analyzer.
4. Carefully remove the entire mounting board with analyzer from the inner box.
NOTE
Do not lift or pull the analyzer by the fluidics or the electronic components.
Use the back panel to lift the analyzer system.
5. Unbolt the analyzer from the mounting board by removing the four mounting bolts with a
9/16’’ wrench. These bolts may be discarded.
6. Carefully place the analyzer at a convenient location until proper installation can be
completed.
2.1
Mounting and Plumbing Instructions
WARNING
Do not connect power prior to the mounting and plumbing of the analyzer.