4-13
IM 04P01B01-17E
Commands
4
SC
Sets the chart speed.
Syntax
SC p1
<terminator>
p1
Chart speed
Query
SC?
Example
Set the chart speed to 25 mm/h.
SC 25
Description Select the chart speed from the list of choices
below.
Pen model
5 to 12000 mm/h (82 levels, see section 4.3)
Dot model
1 to 1500 mm/h (1 mm steps)
SD
Sets the date and time.
Syntax
SD p1
<terminator>
p1
Date and time (fixed to the YY/MM/DD
HH:MM:SS format)
YY
Year (00 to 99, the lower 2
digits)
MM
Month (01 to 12)
DD
Day (01 to 31)
HH
Hour (00 to 23)
MM
Minute (00 to 59)
SS
Second (00 to 59)
Query
SD?
Example
Set the recorder clock to 13:00:00, December 1,
2004.
SD 04/12/01 13:00:00
Description • The format of p1 is fixed to 17 characters. An
error results if a space is inserted in positions
other than the specified positions.
• On the
µ
R10000 with system version 1.02 or
earlier, the parameter p2 can be used.
p2
Standard time or DST (winder, summer)
Winter
Standard time
Summer
DST
VT
Sets the dot printing interval
(dot model).
Syntax
VT p1
<terminator>
p1
Dot printing interval setting.
AUTO
Automatically adjust the dot
printing interval according to
the chart speed.
FIX
Record at the fastest printing
interval.
Query
VT?
Example
Record at the fastest printing interval.
VT FIX
Description When trend recording is performed on 6
channels, the fastest printing interval is 10
seconds.
SZ
Sets zone recording.
Syntax
SZ p1,p2,p3
<terminator>
p1
Channel number
p2
Leftmost position of the zone (
µ
R10000: 0
to 95,
µ
R20000: 0 to 175) [mm]
p3
Rightmost position of the zone (
µ
R10000: 5
to 100,
µ
R20000: 0 to 180) [mm]
Query
SZ[ p1]?
Example
Display channel 02 in a zone between 30 and 50
mm.
SZ 02,30,50
Description • Be sure that p3 is greater than p2 and that the
zone width (p3 – p2) is greater than or equal
to 5 mm.
• On models with the computation function (/M1
option), computation channels can be
specified.
SP
Sets the partial expanded
recording.
Syntax
SP p1,p2,p3,p4
<terminator>
p1
Channel number
p2
Partial expanded recording ON/OFF (ON,
OFF)
p3
Boundary position (1 to 99) [%]
p4
Boundary value
Query
SP[ p1]?
Example
Set the 25% position of channel 01 to 1.000 V.
SP 01,ON,25,1000
Description • When the input range is set to SKIP (SR
command), p2 cannot be turned ON.
• Set p3 as a percentage where 100 mm is
assumed to be 100%.
• Parameter p4 can be set in the range
(leftmost value of span + 1) to (rightmost
value of span – 1). If scaling is enabled, the
range is (leftmost value of scaling – 1) to
(rightmost value of s 1). Set the
parameter using an integer.
• This setting is possible when partial expanded
recording is enabled (UK command).
• This command cannot be used if the partial
expanded recording range does not exist
(when the span width is set to 1, for example).
• On models with the computation function (/M1
option), computation channels can be
specified.
• When the computation equation setting (SO
command) is turned OFF, p2 cannot be
turned ON.
4.4 Setting Commands