IM 12D06A03-01E
Specifications 2-9
・
Electrical data of the ISC202S.
- Supply circuit (ter and -):
Maximum input voltage V
max
= 31.5 V.
Maximum input current I
max
= 100 mA.
Maximum input power P
max
= 1.2 W.
Effective internal capacitance C
i
= 22 nF.
Effective internal inductance L
i
= 35
P
H.
- Sensor input circuit (terminals 11 through 17):
Maximum output voltage V
t
= 14.4 V.
Maximum output current I
t
= 20 mA.
Maximum allowed external capacitance
C
a
= 600 nF
Maximum allowed external inductance
L
a
= 88 mH.
・
If Hand Held Terminal (HHT) is not connected to the power supply lines of the ISC202S (see figure 1):
Any FM Approved barrier or power supply may be used that meets the following requirements.
V
oc
or V
t
d
31.5 V; I
sc
or I
t
d
100 mA; C
a
t
22nF + C
cable
; L
a
t
35
P
H + L
cable
If HHT is connected to the power supply lines of the ISC202S (see figure 2):
The Hand Held Terminal must be FM Approved. Refer to the manufacturers control drawing of the
HHT and the barrier/power supply to determine the cable parameters.
(V
oc
or V
t
) + V
HHT
d
31.5 V; (I
sc
or I
t
) + I
HHT
d
100 mA;
C
a
t
22nF + C
cable
+ C
HHT
; L
a
t
35
P
H + L
cable
+ L
HHT
When installing this equipment, follow the manufacturer ’s installation drawing.
Installation should be in accordance with ANSI/ISA RP 12.06.01 “Installation of Intrinsically Safe
Systems for Hazardous (Classified) Locations” and the National Electrical Code (ANSI/NFPA 70).
Control equipment connected to the barrier/power supply must not use or generate more than 250
Vrms or Vdc.
・
Resistance between Intrinsically Safe Ground and earth ground must be less than 1.0 Ohm.
・
In case of using cable glands in Outdoor location, they shall be UV rated or made of metal.
WARNING
- Substitution of components may impair Intrinsic Safety
- To prevent ignition of flammable or combustible atmospheres, disconnect power before servicing or
read, understand and adhere to the manufacturer ’s’live maintenance procedures.
ISC40S Sensor
terminals 11-17
Max. cablelength: 60 mtr.
Cable dia. : 3…12 mm.
ISC40S Sensor
terminals 11-17
Max. cablelength: 60 mtr.
Cable dia.: 3…12 mm.
Ùnclassified Location
FM Approved
Power Supply
(HART compatible)
Output
Supply
+
_
G
Classified Location
Unclassified Location
+
_
G
Functional
earth
Fanctional
earth
Intrinsically safe design
FM Class I, Div.1, Group ABCD,
T4 for ambient temp. < 55°C
T6 for ambient temp. < 40°C
ISC202S transmitter
+
-
Load
Resistance
Classified Location
Fanctional
earth
+
_
24 volts DC Nominal
Supply Voltage.
FM Approved safety barrier or
power supply
with Rint = 300
:
(HART compatible)
For electrical data:
see text below.
For electrical data:
see text below.
Intrinsically safe design
FM Class I, Div.1, Group ABCD,
T4 for ambient temp. < 55°C
T6 for ambient temp. < 40°C
ISC202S transmitter
Figure 1
Figure 2
2-5. Control Drawing ISC202S mA HART® Specification (FM Intrinsically safe design)
Application Doc. No.: IKE028-A10 P.5 to P.6
Содержание vigilantplant EXA ISC202G
Страница 8: ...IM 12D06A03 01E LT LV PL EST SLO H BG RO M CZ SK...
Страница 12: ...IM 12D06A03 01E 1 4 Introduction...
Страница 53: ...IM 12D06A03 01E Parameter setting 5 11 5 2 4 Temperature compensation for first conductivity value To TEMP 2...
Страница 54: ...IM 12D06A03 01E 5 12 Parameter setting 5 2 5 Temperature Compensation for second conductivity value To SERV...
Страница 68: ...IM 12D06A03 01E Spare parts 10 1 10 SPARE PARTS See Customer Maintenance Parts List...
Страница 92: ...IM 12D06A03 01E 2 20 Appendix...
Страница 97: ...IM 12D06A03 01E Appendix 3 5...
Страница 102: ...IM 12D06A03 01E 3 10 Appendix...
Страница 105: ...IM 12D06A03 01E Appendix 3 13 3 3 QIS 12D06A03 61E Figure 1 Testing Circuit and Test Equipment...
Страница 106: ...IM 12D06A03 01E 3 14 Appendix...
Страница 110: ...IM 12D06A03 01E 3 18 Appendix...