<4. OPERATION>
4-9
IM 12T03A01-02E
4.3.6
Pressure Compensation
Pressure compensation means to compensate the shift generated by measurement under
operating pressure. The compensation should be carried out as necessary only for calibration
under atmospheric pressure.
Measured value is affected up to +0.00005 g/cm
3
by 100 kPa pressure variance.
Therefore, for liquid under pressure approximating to the atmospheric pressure calibrated
value under atmospheric pressure can be used. Pressure compensation is to be carried out to
minimize effects generated by high operating pressure.
Compensation data is obtained by conducting sample liquid. Measure the data according to the
following procedure when the liquid density is stable. The procedure assumes that sample liquid
is already conducted according to Section 4.3.3.
(1) Fully open the needle valve “NV4” and read P from the pressure gauge. At the same time,
obtain density at the reference temperature indicated on the density converter.
(2) Fully close the ball valve “BV1”, and open the needle valve “NV3” to equalize the sample
liquid pressure in the detector to the atmospheric pressure. Obtain density at the reference
temperature indicated on the density converter.
(3) Calculate the compensation data using the densities obtained in (1) and (2).
Compensation data (g/cm
3
) =
Density under working pressure – density under atmospheric pressure
When calibration the compensation data should be subtracted from the density value of the
standard solution.
STEAM IN
CONDENSATE OUT
V2
a
a
V1
b
ST1
a'
a'
b'
TI
F
FM
NV1
PI
NV2
DD
NV3
NV4
BV1
Element specifications
F:
Strainer body; 316 SS
Element; 316 SS, or Ni (option)
PI: Pressure gauge, 0 to 1MPa or 0 to 2MPa, 316 SS
TI: Thermometer. 0 to 100 °C or 0 to 150 °C, 316 SS
FM: Flowmeter, tapered metal tube flowmeter, 1 to 10 L/min, 316 SS
BV1: Ball valve, 316 SS
NV1 to 4: Needle valve, 316 SS
DD: Density detector
V1, V2: Valve
ST1: Steam trap
SAMPLE IN
SAMPLE OUT
Figure 4.5
Flow Diagram
5th Edition : Dec. 11, 2015-00
Содержание vigilantplant DM8
Страница 1: ...User s Manual R Model DM8 Vibration Type Density Meter IM 12T03A01 02E IM 12T03A01 02E 5th Edition ...
Страница 5: ...Blank Page ...
Страница 9: ...Blank Page ...
Страница 25: ...Blank Page ...
Страница 41: ...Blank Page ...
Страница 57: ...Blank Page ...
Страница 63: ...Blank Page ...