<8. TROUBLESHOOTING>
8-1
IM 12E04A02-02E
8. TROUBLESHOOTING
If a failure occurs in the converter or the detector, the self-diagnostic function detectors the failure
and lights up FAIL lamp to notify of the occurrence of the failure. When the FAIL lamp is lit,
fi
rst
check the error code and then do the checks for each item described in Section 8.1.
Incorrect indications that cannot be checked by the self-diagnostic function may also be
considered. If an indication failure occurs but the FAIL lamp is not lit, do the checks referring
to each item in Section 8.2. For checking, a multimeter, or the like, that can measure DC/AC
voltage, DC currents, and resistances, will be needed.
As a result, if component replacement or repair is required, or if the location of the failure cannot
be determined, contact Yokogawa.
8.1 Events in Which FAIL Lamp Is Lit
If the FAIL lamp is lit during measurement, check the error code in the following procedure and
carry out making checks corresponding to each number.
<Error Code Con
fi
rmation Procedure>
1. Select FUNCTION “A” in the <MAINT> mode and press the [ENT] key.
2. A display appears, such as “Err11”. Check the points according to the listed procedure for
each error number.
(Note 1) When a failure occurs, the FAIL contact operates. If the FAIL status presents an
inconvenience, perform an “error reset operation” (FUNCTION “B” in the <MAINT.>
mode) to reset the failure. If resetting, be sure to
fi
rst check the details by the “error
code checking operation” before resetting.
(1) Err11 (turbidity overrange)
Carry out checking in the following order.
Operation and checking procedure
Failure Mode and Corrective Action
1. Measure the DC voltage across the converter
connection terminals [S1(+) and AG(-)] to make
sure of it, and con
fi
rm that it is smaller than -2.45 V
or larger than +0.5 V.
2. Check whether or not the turbidity of measuring
water exceeds the measurable range (2200 mg/l
(deg.)).
a. “Err11” is detected when the turbidity of the
measuring water increases and the detector signal
exceeds -2.45 V.
3. Measure the DC voltage across the converter
connection terminals [V+(+) and AG(-)] and con
fi
rm
that it is +5 ±0.5 V.
b. If the measured voltage is out of the range, a
failure of the +5 V power circuit on the converter
analog board is assumed.
=> Request inspection and/or repair of the
analog board.
4. Measure the DC voltage across the converter
connection terminals [V-(+) and AG(-)] and con
fi
rm
that it is -5 ±0.5 V.
c. If the measured voltage is out of the range, a
failure of the -5 V power circuit on the converter
analog board is assumed.
=> Request inspection and/or repair of the
analog board.
5. Introduce the external light onto the scatteredlight-
receiving element by opening the detector and
check whether the DC voltage across the converter
connection terminals [S1(+) and AG(-)] changes.
d. If the measured voltage does not change, a failure
of the detector preampli
fi
er is assumed.
=> Request inspection and/or repair of the
measuring preampli
fi
er.