<6. PROCEDURES FOR CONVERTER OPERATION>
6-32
IM 12E04A02-02E
(28) Selecting Execution/Stop of Spike Detection Function (FUNCTION
"F.")
(Example of display)
1. Select FUNCTION "F." in the <PROGRAM1>
=> OFF (stop)
mode. The current set status is displayed.
2. Press the [>] key to select on/off.
=> ON (execution)
F6601.ai
ON
OFF
3. Press the [ENT] key to enter the setting.
=> ON
The
display
fl
ashes once.
6.7 Operation in <PROGRAM2> Mode
In this mode, the time setting operations for automatic cleaning and automatic zero calibration,
the operation settings for contact output and the selection operations for part of execution/stop of
the failure detection function are curried out.
• FUNCTION "1" to FUNCTION "F": Time settings, contact output operation settings
• FUNCTION "1." To FUNCTION "F.": Selection for part of execution/stop of the failure
detection functions
Key operations made by specifying a function in the <PROGRAM2> mode and the displays
following those key operations are described in the following order.
(1) Time setting for automatic cleaning and automatic zero calibration
(2) Selection of contact output operation (open/close)
(3) Selection of execution/stop of the failure detection functions
(1) Time Setting for Automatic Cleaning and Automatic Zero Calibration
(FUNCTION "1" to FUNCTION "A")
Table 6.12 shows a list of setting times and operation functions. For details on automatic cleaning
and automatic zero-calibration operations, see subsections 4.2.2 and 4.2.3. For speci
fi
cation
without automatic cleaning or automatic zero calibration (Suf
fi
x codes: -NN, -A1), this setting is
non-operative.
Table 6.12 Time Settings for Automatic Cleaning and Automatic Zero Calibration
Setting
FUNCTION
Setting range
Factory setting
Automatic cleaning period
1
0.1 to 24.0 hours
0002.0
Cleaning calibration ratio *3
2
1 to 10 times
00001
Repeated cleaning
3
1 to 20 times
00005
Cleaning time 1 *1
4
10 to 120 seconds
00030
Cleaning time 2 *2
5
10 to 120 seconds
00030
Draining time 1 *1
6
10 to 120 seconds
00010
Draining time 2 *2
7
10 to 120 seconds
00010
Filling time 1 *1
8
10 to 120 seconds
00100
Filling time 2 *2
9
10 to 120 seconds
00100
Relaxation time
A
30 to 600 seconds
00150
(*1)
For models with Suf
fi
x code -A2 (without automatic zero calibration), use this time setting.
(*2)
For the models with Suf
fi
x code -A3 (with automatic zero calibration), use this time setting.
(*3)
The time is set for the automatic cleaning period, while for the automatic zero-calibration period, the cleaning calibration ratio,
which means the number of the automatic cleaning is performed for one automatic zero calibration, is set.
3rd Edition : May. 31, 2010-00