<8. Maintenance>
8-7
IM 12K01B02-01E
8.2.4
Replacement of Desiccant Agent in the Measuring Cell
Here describes the replacement of desiccant agent in the measuring cell of detector.
Replace the desiccant agent every 6 months.
1. Turn ON the maintenance switch.
2. Turn OFF the power switch.
3. Remove the detector from the measuring tank.
4. Lightly wipe off moisture adhered to the measuring cell with soft cloth.
5. Remote the 2 brackets from the detector.
The following description is the replacement method for the desiccant agent of the light
source cell.
Replace the desiccant agent on the detector side by following the same procedure.
Bracket
6. Remove the screw.
Screw
7. Remove the cell plate D.
Cell plate D
8. Remove the cell packing by using tweezers.
Cell plate C
Cell packing
Tweezers