3 Setting Up with the Personal Computer
IM 54011-E
■
Setting logs
●
Make settings using a tag/log table
A variety of tag settings can be set beforehand in a table, and from the table, you can select
settings to be downloaded to a TM20 as appropriate.
1.
Run Thermo-Collector.
After the Thermo-Collector main window opens, a document window entitled “Setup Tag/Logs - New” opens in it.
2.
From the File menu, choose Tag/Log Table then New to open the Tag/Log Table window
(a document window).
Tag/Log Table window
3.
Enter the log names in the Abbr. Name column with a maximum of eight alphanumeric
characters (for example, Convey).
4.
In the Full Name column, enter more descriptive tag names.
There is no limitation in length with full names. If you select this table as the reference tag/log table, Thermo-
Collector will replace the abbreviated names with the corresponding full names based on this table when receiving
measured data from a TM20.
5.
Double-click in the “Channel” column.
The specified log name is set and the pull-down menu appears.
6.
Select either A, B, D, or AB.
The specified channel is set.
7.
Click in the “Measuring interval” column.
The cursor moves to a cell for the measurement interval.
8.
Enter the measurement interval with numeric characters and colons “:” (Example:
00:10:00).