Thermo-Collector specifications
●
Features
Measuring ranges
Thermocouple types K, J, E, and T, and
±
100 mV and
±
1 V
Input channels
2 channels (A and B) or single channel (D)
Potential difference between channels
Must not exceed
±
1 V (when the RS-232-C cable is not connected).
Measured temperature ranges and resolutions
For input channels A and B at a TM20 (not at a PC)
Range setting
Measured temperature range
Resolution
Thermocouple
type K
–200.0 to 1372.0
°
C
0.1
°
C
type J
–200.0 to 1000.0
°
C
0.1
°
C
type E
–200.0 to 700.0
°
C
0.1
°
C
type T
–200.0 to 400.0
°
C
0.1
°
C
Voltage input
100 mV
–100.0 to 100.0 mV
0.1 mV
1 V
–1.000 to 1.000 V
0.001 V
Channel D — Non-contact thermo probe (Model 900 03)
• Measured temperature range: –20.0 to 400.0
°
C
• Resolution: 1
°
C
Accuracy
Input channels A and B:
(under the ambient temperature of 23
±
5
°
C and relative humidity of 20 — 80%, when the low battery alarm is not
occurring, and excluding the accuracy of reference junction compensation)
Input type
Range
Accuracy
Thermocouple
≥
–100
°
C
±
(0.1% of r 0.3
°
C)
< –100
°
C
±
(0.1% of r 0.6
°
C)
Reference junction compensation
—
±
0.4
°
C (when the temperature at input jacks are stabilized)
Voltage
100 mV or 1 V
±
(0.1% of r 0.2% of range)
Accuracy (error of reading) of input from non-contact thermo probe (channel D):
The error of reading of the input from a non-contact thermo probe that can be connected to the TM20 (i.e., Model
900 03) is
±
1
°
C excluding the error of the probe itself. (For the error of the non-contact thermo probe, see the
respective specifications.)
Effects of ambient conditions
Effects of ambient temperature changes on thermocouple inputs (channels A and B):
±
(0.01% of r 0.03
°
C)/
°
C or less
Effects of ambient temperature changes on reference junction compensation:
±
0.03
°
C/
°
C or less (at ambient temperature of 0 – 50
°
C)
Effects of ambient temperature changes on voltage inputs (channels A and B):
±
(0.01% of r 0.02% of range)/
°
C or less
Effects of humidity changes (at relative humidity of 20 – 80%):
±
1 digit or less
Battery life
(when not using the backlight of the LCD)
When using collector function:
Approx. 2 days (under continuous 24-hour operation) (when the RS-232-C
cable is not connected.)
Approx. 5 days (8 hours of measurements and 30 minutes of communication
every day)
When using logging function:
Approx. 2 days (measurement intervals < 10 seconds) (when the RS-232-C
cable is not connected.)