<APPENDIX>
App.-1
IM 12E01B30-02EN
1st Edition : Mar. 08, 2023-00
Appendix Automatic cleaning/Automatic
zero calibration Sequence
n
Automatic cleaning (-A2, -A5)
CLOSE
OPEN
CLOSE
OPEN
Washing
Washing
Meas.
Status
SV2(Wash)
Drainage time
Drainage time
SV4(Drain)
Recovery Meas.
Flow/Calibration
Waiting time
Flow/Calibration
Waiting time
Recovery
1: When Automatic cleaning starts, SV4 (Drain) opens and starts draining the sample out of
the Flow cell.
2: After the “Drainage time” elapses, SV4 closes and sample starts flowing into the detector.
3: After the “Flow/Calibration Waiting time” elapses, SV2 (Wash) opens. Cleaning water (tap
water) starts purging from two water ports on the side of the Flow cell to wash the side or
edge of the Flow cell
4: After the cleaning time elapses, SV4 opens during the Drainage time to wash the Flow cell,
head tank, and the pipes between head tank and the detector.
5: If “Repeat count of wash” is set to over 2, valve working 2 to 4 is repeated.
6: Once the draining for cleaning completes, SV4 closes, and TB830D starts feeding a
sample.
7: After Flow/Calibration Waiting time” elapses, the detector becomes Recovery time.
8: Afthe the Recovery time elapses, a measurement starts.
n
Automatic cleaning (-A3, -A5)
CLOSE
OPEN
CLOSE
OPEN
CLOSE
OPEN
CLOSE
OPEN
Washing
Flow/Calibration
Waiting time Recovery
Washing
Recovery Meas.
Drainage time
Drainage time
Drainage time
Washing
Flow/Calibration
Waiting time
Flow/Calibration
Waiting time
Meas.
Status
SV1(Measure)
SV2(Wash)
SV3(Zero)
SV4(Drain)
1: When Automatic cleaning starts, SV1 (Measure) closes and SV4 (Drain) opens to start
draining the sample out of the Flow cell.
2: After the “Drainage time” elapses, SV4 closes and SV3 (Zero) opens to feed the zero water
into the detector.
3: After the “Flow/Calibration Waiting time” elapses, SV2 (wash) opens. Cleaning water (tap
water) starts purging from two water ports on the side of the Flow cell to wash the side or
edge of the Flow cell.