<9. TROUBLESHOOTING>
9-2
IM 12E04B40-02EN
1st Edition : Mar. 08, 2023-00
Alarm
Number
Name
Explanation
Remedy
NE107
Default
(*1)
NE107
Changed
X54E
Auto Zero cal result
error
Result of auto zero
calibration is out of
range.
Confirm the solenoid valve
runs correctly by manual
operation. Check if the
zero water flows normally.
If no error is found in the
sampling, check whether no
condensation is found on
Measurement window.
S
Enable
X54F
Auto Cal stability time
limit
Detection value is
not stable during
auto calibration.
S
Enable
Note : X of Alarm Number denotes Channel of the measurement detection unit.
1: sensor connection number 1-1
5: sensor connection number 2-1
6: sensor connection number 2-2 (not required on the first FD)
(*1): F: Failure, C: Function Check, S: Out of Specification, M: Maintenance required, N: Off
Figure 9.2
TB830D Analyzer Error Message List (Wash)
Alarm
Number
Name
Explanation
Remedy
NE107
Default
(*1)
Error
config.
(HMI)
009E
Auto wash config.
error
Automatic cleaning
is not properly set.
Check automatic cleaning
setting.
Converter setting > Auto wash
> Auto calib. setting
C
Enable
009F
Disable auto update of
next start
Automatic cleaning/
Automatic zero
calibration is
overdue.
Set the first calibration start
time. Converter setting > Auto
wash> Auto calib. setting
C
Enable
(*1): F: Failure, C: Function Check, S: Out of Specification, M: Maintenance required, N: Off
9.2
When No Error Indication Appears
There is a possibility that measurement has abnormalities without any error occurrence. Next
table shows some possible causes. Solve the problems referring to the next table.
Figure 9.3
Abnormalities, Their Possible Causes, and Remedies
Abnormalities
Possible Causes
Remedies
Measured value
fluctuates largely.
● Sample air bubble or contamination
stays in Flow cell.
● The measured value fluctuates
according to the sample turbidity.
●
Clean the Flow cell.
●
Check the head tank before the sample
inlet of the detector.
●
Check if the turbidity reading is stable
by using Zero water or Check tool. If the
reading is not stable, identify where the
error occurs in Flow cell, Light source, or
Detector board. After then, bring it to a
normal state.
●
If no error is found in the instrument
or sample flow, set larger value of
measurement time constant.
Measured value
increased over time
● Inside of the Flow cell is contaminated.
● Water droplets or contamination is
adhering to the lenses of the light
source or detection unit.
●
Check the condition of Flow cell,
head tank, piping, lenses. If they are
contaminated, clean them.
●
Check the condition of desiccant
attached inner the cover. Replace it with
new one if deterioration is found.
Measured value
abruptly increases
but plunged in a
short time
● Air bubbles or contamination pass
through the Flow cell.
●
Check the condition of Flow cell,
head tank, piping, lenses. If they are
contaminated, clean them.
●
Turn ON the Air denoising function.
●
Optimize the air denoising parameter.