<9. TROUBLESHOOTING>
9-1
IM 12E04B40-02EN
1st Edition : Mar. 08, 2023-00
9. TROUBLESHOOTING
9.1
When an error occurs
Figure 9.1
TB830D Error List
Alarm
Number
Name
Explanation
Remedy
NE107
Default
(*1)
NE107
Changed
X500
EEPROM error
The TB830D has become defective. Contact
Yokogawa service.
F
Disable
X501
User param. read error
F
Disable
X502
Factory param. read error
F
Disable
X503
AD converter failure
F
Disable
X504
RAM failure
F
Disable
X505
Flash Memory failure
F
Disable
X50A
Sensor communication
error
No sensor is
detected.
Check the power supply
connection.
Check the connection between
the detection unit and the
converter.
F
Enable
X520
Wash/Cal. box error
Cleaning box does
not work.
Check the connection of Relay
box or configuration (Enable/
Disable.) of the cleaning box.
S
Disable
X521
Humidity sensor failure Humidity sensor
does not work.
After turning ON/OFF the power,
if the problem still remains,
check the cable connection on
the detector board.
If the cable is correctly
connected, replace the detector
board. When you replace the
detector board, you need to
perform turbidity calibration.
S
Enable
X522
Lens heater failure
Lens heater shows
abnormal resistivity.
Check the condition of the anti-
fog lens heater.
Check for water leaks to the
lens and disconnection of the
lens heater drive cable. If there
are no abnormalities in these,
the lens heater may have
reached the end of its life.
Contact Yokogawa.
S
X540
Turbidity High
Turbidity value is
exceeded high limit
setpoint.
Check measurement value.
Confirm Turbidity warning high
limit on sensor setting.
S
Enable
X541
Turbidity Low
Turbidity value is
less than low limit
setpoint.
Check measurement value.
Confirm Turbidity warning low
limit on Sensor setting.
S
Enable
X542
Light detection error
Photodetector
current is out of
range.
Check the scattered light current.
Identify where the error occurs
: Flow cell, light source board,
detector board. After then, bring
it back to a normal state.
S
Enable
X543
Light source failure
Light source intensity
is abnormal.
Check light source monitoring
current and Lightning IF. If value
is abnormal, open the front cover
and check the cable connection
or assemble condition. If the
cause is not found, please
replace the Light source part.
S
Enable
Note : X of Alarm Number denotes Channel of the measurement detection unit.
1: sensor connection number 1-1
5: sensor connection number 2-1
6: sensor connection number 2-2 (not required on the first FD)
(*1): F: Failure, C: Function Check, S: Out of Specification, M: Maintenance required, N: Off
Содержание TB830D
Страница 7: ...Blank Page ...
Страница 29: ...Blank Page ...
Страница 35: ...Blank Page ...
Страница 51: ...Blank Page ...
Страница 75: ...Blank Page ...
Страница 85: ...Blank Page ...