<8. MAINTENANCE>
8-1
IM 12B01B30-02EN
1st Edition : Mar. 08, 2023-00
8. MAINTENANCE
This chapter describes the inspection and maintenance required to keep the instrument in good
operating condition.
n
Inspection/Maintenance Items and Intervals
The main inspection/maintenance items and their recommended intervals to keep the instrument
in good operating condition are provided in Table 8.1.
Since the frequency of inspection/maintenance depends on the operating conditions, use the
recommended intervals shown in Table 8.1 as a reference and perform inspection/maintenance
at appropriate intervals.
Table 8.1
Inspection/Maintenance Items and Recommended Intervals
Inspection/Maintenance Item
Recommended Interval Reference
section
Cleaning with tap water
As needed
8.1
Cleaning of Flow cell
Weekly
8.2
Cleaning of head tank
Monthly
8.3
Checking and replacing desiccants
6 months
8.4
Pipe cleaning
6 months
8.5
Lens cleaning
6 months
8.9
Inspection solenoid valve
6 months
8.6
Replacement of zero turbidity filter element
Yearly *1
8.7
LED replacement
3 Years
8.8
Fuse replacement
As needed
8.10
Replacement of the Silicone Tube *2
Yearly
8.11
Checking flow rate (liquid level)
Monthly
8.12
Zero calibration
Monthly
7.4, 7.5
Slope calibration
Monthly
7.6
,
7.7
*1:
This is an interval estimated if tap water used has average turbidity of 0.4 dgrees and runs at a flow rate of 2 L/min.
*2:
With sampling system (-A1, -A2, -A3)
8.1
Cleaning with cleaning water
When Without automatic cleaning (Sampling system:-NN,-A1) is selected, the valve is manually
opened/closed and cleaning is performed.
Wash while running the zero water to avoid splashing the cleaning water.
(1) Set converter maintenance mode to On.
(2) Loosen screw A at the bottom of the detector and open the top of the detector.
(3) Close the valve V1 and open V3 to allow zero water flow.
Open valve V4 gradually to prevent splashing of cleaning water from the Flow cell. With the
valve V4 open, wash for approximately 30 seconds.
(4) Close the valve V4 and open the valve V5 for about 10 seconds to drain the water.
(5) Repeat steps (3) and (4) several times, depending on stain severity.
(6) Close the valve V5.
(7) Close the upper part of the detector and tighten the screw A.
Set the valve to the state at the time of measurement and turn Off the maintenance mode.
Automatic cleaning is performed periodically when With sampling system (Sampling system:-A2,
-A3) is specified. See “4.5 Automatic cleaning/Automatic zero calibration setting”
for automatic
cleaning/automatic zero calibration.
When (Sampling system :-A1) is specified: Cleaning is performed by manually operating the
valve V4 for a wash.