< 7. Calibration >
7-9
IM 12E01B30-02EN
1st Edition : Mar. 08, 2023-00
7.10 Sensitivity correction
Use Sensitivity correction to adjust a difference sensitivity between a standard solution and a
grab sample after both Zero calibration and Slope calibration are conducted correctly. Calculate
the sample turbidity manually and use the value as a calibration value.
(1) Turn on the maintenance mode.
(2) Press on Main screen and go to the calibration menu.
(3) On the calibration menu, select [Sensitivity correction] in the menu list of Calibration others.
(4) Run or feed from top of the detector a sample to use for the Sensitivity correction.
(5) Confirm the reading stable. Press [check stability].
(6) After the message pops saying the reading is stable, enter the calibration value (turbidity
value to adjust). Then press [Next].
(7) Check the result of calibration and press [Accept] to update the Sensitivity correction factor
of the calibration parameter.
(8) Drain the sample and clean the Flow cell with zero turbidity water. Place the clear cover and
the top cover on top of the detector. If you finish the calibration, run the sample and turn off
the maintenance mode.
7.11 2 points correction
The Zero point and Sensitivity correction factor are adjusted. Use two arbitrary known turbidity
solution. Take two measurements and record the low concentration as low point, the high
concentration as high point. Be sure to take low point first, then do the measurement of the high
point.
[How to perform a two point calibration]
(1) Turn on the maintenance mode.
(2) Feed the solution for Low point measurement into the Flow cell.
(3) Press on Main screen to go to the calibration.
(4) Select [2point calibration] in the menu list of Calibration others.
(5) Confirm the reading stable. Press [check stability].
(6) When a message pops saying that the reading is now stable, enter the calibration value.
Press [Next].
(7) Feed the solution for High point measurement into the Flow cell. Wait until the reading
becomes stable. After it becomes stable, press [check stability].
(8) When a message pops saying the reading is now stable, enter the calibration value. Press
[Next].
(9) Confirm the result of calibration. Press [Accept] to update the parameter for Zero shift value
and sensitivity correction factor.
(10) Drain the sample and clean the Flow cell with zero turbidity water. Place the clear cover and
the top cover on top of the detector. If you finish the calibration, run the sample and turn off
the maintenance mode.