< 4. SETTING PARAMETERS >
4-6
IM 12E04B40-02EN
1st Edition : Mar. 08, 2023-00
4.2
Cal./Maint. Settings
4.2.1
Turbidity value of check tool
Set up the turbidity of the check tool used for slope calibration. The value is set at the factory.
CAUTION
When the initialization is executed, setup data at the factory is deleted. After initialization, re-enter
the turbidity value of the check tool.
4.2.2
Zero shift value, Correction factor
These parameters are related to zero shift correction and sensitivity correction in the calibration
menu. Normally zero shift value is 0, and the correction factor is 1. See “7. CALIBRATION”.
4.2.3
Stability check
You can configure the following parameters to evaluate the measurement stability in a calibration.
• Stability check allowable width
• Stability checking time
• Stability check limit time
The measurement stability is verified, when the measurement stays within the range of “Stability
check allowable width” without exceeding the range for longer than the set period of” Stability
checking time”. If the measurement does not become stable before “Stability check limit time” is
up, a confirmation dialog appears to abort the calibration.
4.2.4
Wash/Cal. box
Select whether or not to use Wash Box (Relay box for solenoid valve). While using the Relay box,
make the Wash box Enabled.
If Wash Box is disabled, you cannot implement automatic cleaning/automatic zero calibration.
If you enable this function when Wash Box is not connected, an alarm occurs to notice the Wash
Box error.
4.2.5
Calibration setting others
Zero, slope, Sens.factor Linearize table changed manually. Go to the sensor detail screen to
check the setting as a reference. These parameters are important for the turbidity linearity.
Do not change those setting except for resetting values.
4.3
Diagnosis setting
4.3.1
Light source diagnosis
Select Light source diagnosis On/Off. See 5.3 for details.
4.3.2
Dried condition diagnosis
Select Dried condition diagnosis On/Off. See 5.3 for details.