<2. INSTALLATION, PIPING, AND WIRING>
2-7
IM 12B01B30-02EN
1st Edition : Mar. 08, 2023-00
L
N
G
Power
(201) L
(202) N
(002) G
Power
Converter
SENCOM
L (101)
N (102)
G (001)
L (101)
N (102)
G (001)
RS485 (-R)
Ethernet (-E)
+
–
+
–
+
–
+
–
+
–
(61)
(62)
mA1
(65)
(66)
mA2
(81)
(82)
mA3
(85)
(86)
mA4
(21)
(22)
(24)
SHLD (63)
(87)
(88)
SHLD (89)
A+ (91)
B
−
(92)
GND (93)
SHLD (94)
Termination (95)
Sensor1
Sensor2
(32) NC
(31) C
(33) NO
S1
(42) NC
(41) C
(43) NO
S2
(52) NC
(51) C
(53) NO
S3
(72) NO
(71) C
(73) NC
S4
Power
Supply
Power
Supply
Power
Supply
Analog output (-N2, -N4)
Analog output (-N2, -N4)
Analog output (-N4)
Analog output (-N4)
Contact input (-N2, -N4)
Analog input (-N4)
Contact output
(-WR)
TB830D
Surface scattering turbidity detector
FLXA402T
Liquid Analyzer for Turbidity and Chlorine
L (101)
N (102)
G (001)
L (201)
N (202)
G (003)
L (101)
N (102)
G (001)
Terminal box
B
L (101)
N (102)
G (001)
L (201)
N (202)
G (003)
L (101)
N (102)
G (001)
CP
AR
Terminal box A
Power supply cable(E)
Power supply cable
TB830D power supply
cable(C)
Power supply cable(E)
FLXA402T power supply cable (B)
SENCOM cable for Turbidity/Chlorine (A)
• If a Relay box is not provided, wire power cable (E) to TB830D.
• If Terminal box A or B is included, wire the power supply cable (E) to the Terminal box.
• wire TB830D power supply cabling (C) between the junction box and TB830D.
Figure 2.6
Wiring TB830D-□-□-NN (without sampling system)
L
N
G
Power
Sensor1
Sensor2
(32)NC
(31)C
(33)NO
S1
(42)NC
(41)C
(43)NO
S2
(52)NC
(51)C
(53)NO
S3
(72)NO
(71)C
(73)NC
S4
RS485(-R)
Ethernet(-E)
Power
Converter
SENCOM
L (101)
N (102)
G (001)
L (201)
N (202)
G (003)
(201) L
(202) N
(002) G
L (101)
N (102)
G (001)
+
–
+
–
+
–
+
–
+
–
(61)
(62)
mA1
(65)
(66)
mA2
(81)
(82)
mA3
(85)
(86)
mA4
(21)
(22)
(24)
SHLD (63)
(87)
(88)
SHLD (89)
A+ (91)
B
−
(92)
GND (93)
SHLD (94)
Termination (95)
FLXA402T
Liquid Analyzer for Turbidity and Chlorine
Power
Supply
Terminal box B
TB830D
Surface Scattered Light Turbidity Detector
Analog output (-N2, -N4)
Analog output (-N2, -N4)
Analog output (-N4)
Analog output (-N4)
Contact input
(-N2, -N4)
Analog input (-N4)
Contact output
(-WR)
CP
AR
L (201)
N (202)
G (003)
L (101)
N (102)
G (001)
Power
Supply
Terminal box A
L (101)
N (102)
G (001)
Power supply cable (E)
Power supply cable (E)
• Relay box A or B is included.
• wire power supply cable (E) to the junction box.
Figure 2.7
Wiring diagram with sampling system (for-A1)