<2. INSTALLATION, PIPING, AND WIRING>
2-2
IM 12B01B30-02EN
1st Edition : Mar. 08, 2023-00
n
Installation
See GS12E04B40-01EN for the dimensions of detectors and mounting brackets. For the
converter, see the user’s manual of FLXA402T IM 12A01G01-02EN.
It the sampling system is not provided, it can be wall mounted or pipe mounted.
For wall mounting (Option: / R), mount with 4 M5 screws (not included).
For pipe mounting (Option code: / U), mounting bracket for replacement is available.
See GS 12E04B40-01EN for details.
If the sampling device is provided, fix it to a well-drained concrete foundation with anchor bolts
(M12).
530
Approx. 500
Approx. 500
Unit: mm
Approx. 500
Approx. 500
550
100
350
Maintenance space
Maintenance space
Maintenance space
Maintenance space
4-ø15 holes
(for M12 Anchor bolts)
Stand
(100)
(20)
20
490
2.1
Piping
2.1.1
For TB830D-□-□-NN, -A5 (without sampling system)
(1) Sample piping
If the flow rate of the sample is 1.5 to 2 L/min, the sample can be introduced to the detector
as is. Connect the supplied ø33 × ø25 mm black soft vinyl chloride tube to the sample inlet.
However, if the flow rate of the sample exceeds the range of 1.5 to 2 L/min, install a head
tank (constant water level tank) that also serves as a defoaming tank to meet flow rate
requirements. Connect the ø33 × ø25 mm black soft vinyl chloride tube to the pipe between
the head tank and detector.
(2) Drain piping
Connect the supplied ø33 × ø25 mm black soft vinyl chloride tube to the drain port of the
detector.
At this time, install piping with care to prevent accumulation in the tube. Accumulated water
in the drainage pipes may cause problems such as overflow from the detector.
(3) Cleaning water piping
The cleaning water inlet of the detector is sealed with a blind plug.
(Note) When measuring low turbidity (200 NTU or less), prepare a sampling system capable of supplying zero water (filtered tap
water with a zero turbidity filter).