<8. MAINTENANCE>
8-2
IM 12B01B30-02EN 2nd Edition : June 09 , 2021-00
8.1
Washing the Flow cell
The top cover (rubber cover) and the clear cover can be removed to clean the inside of the Flow
cell and the surface of the window glass. If dirt remains, you can remove the sample outlet pipe
and the upper part of the Flow cell to clean every corner in the Flow cell and the window glass.
The Flow cell and window should be washed with mild detergent.
(1) Turn on the maintenance mode.
(2) Draining the sample.
1. Close the sample supply valve.
2. Open the drain valve on the detector to drain the sample.
(3) Washing the Flow cell.
1. Open the zero water supply valve to supply zero water to the detector.
2. Close the drain valve. When the Flow cell is filled with the zero water, open the drain
valve to drain.
3. Repeat steps 1 and 2 at least three times.
(4) Leave the drain valve open.
(5) Remove the top cover. Loosen the four screws that secure the clear cover. and remove the
clear cover. Loosen the four screws that secure the Flow cell and remove the upper part of
the Flow cell. (See Figure 8.1)
Top cover
(rubber cover)
clear cover
(transparent)
Screws
fixing the upper part of
Flow cell (4 points)
Screws fixing
clear cover
(4 points)
Figure 8.1
Removing Top cover, Clear cover, upper part of Flow cell
(6) Thoroughly wash the inside of the Flow cell using a sponge or soft cloth soaked in mild
detergent. Next, carefully wash the 3 glass windows using a brush or a sponge or soft cloth
with cream cleanser. Rinse the inside with zero water (or tap water), as needed.
Window glass
Figure 8.2
Inside the Flow cell and the glass window
(7) Install the upper part of the Flow cell. Fasten the four screws. Be sure to attach the included
gasket (67 mm x 119 mm). If the gasket is deteriorated, arrange the replacement (part
number: K8003NQ). Install the clear cover. Fasten the four screws. Be sure to attach
the included gasket (65 mm x 65 mm). If it is deteriorated, arrange the replacement (part
number: K8003NR).