< 7. Calibration >
7-9
IM 12E01B30-02EN 2nd Edition : June 09 , 2021-00
7.6
Auto zero calibration
Auto wash and auto calibration run one time. Execute this operation in a measurement mode.
Maintenance mode cannot carry out this operation. This operation does not affect any automatic
sequence schedule.
7.7
Zero shift correction
Perform this calibration when you need to adjust the Zero value after both Zero and Slope
calibration are conducted correctly.
1 Turn on the maintenance mode.
2 Press on Main screen and go to the calibration menu.
3 On the calibration menu, select [Zero shift correction] in the menu list of Calibration others.
4 Run or feed from top of the detector a sample to use the Zero shift correction
5 Confirm the reading is stable. Press [check stability].
6 After the message pops saying the reading is stable, enter the value in [Cal.point]. Then
press [Next].
7 Check the result of calibration and press [Accept] to update the Zero shift value of the
calibration parameter.
8 Place the clear cover and the top cover on top of the detector. If you finish the calibration,
run the sample and turn off the maintenance mode.
7.8
Sensitivity correction
Use Sensitivity correction to adjust a difference sensitivity between a standard solution and a
grab sample after both Zero calibration and Slope calibration are conducted correctly. Calculate
the sample turbidity manually and use the value as a calibration value.
1 Turn on the maintenance mode.
2 Press on Main screen and go to the calibration menu.
3 On the calibration menu, select [Sensitivity correction] in the menu list of Calibration others.
4 Run or feed from top of the detector a sample to use for the Sensitivity correction.
5 Confirm the reading stable. Press [check stability].
6 After the message pops saying the reading is stable, enter the calibration value (turbidity
value to adjust). Then press [Next].
7 Check the result of calibration and press [Accept] to update the Sensitivity correction factor
of the calibration parameter.
8 Drain the sample and clean the Flow cell with zero turbidity water. Place the clear cover and
the top cover on top of the detector. If you finish the calibration, run the sample and turn off
the maintenance mode.