<6. OPERATION>
6-1
IM 12E01B30-02EN 2nd Edition : June 09 , 2021-00
6. OPERATION
The TB820D does not normally require any manual operations except when periodic cleaning/
calibration is performed or when a failure occurs.
6.1
Turbidity measurement
When Power Supply is turned ON, the turbidity detector is activated in a measurement mode.
Turbidity of sample is continuously measured. If ultrasonic cleaning is available, ultrasonic is
turned on when power supply is turned on.
6.2
Air denoising
If air denoising function is set enable, measurement value of detector is hold while air noise is
detected. “Check” status icon appears on main screen during air denoising hold time. Contact
output doesn’t work by this function.
6.3
When Sample Water Supply is Cut Off
The turbidity measurement is not affected by temporary suspension of sample water supply,
but long-term suspension may interfere with the accuracy of the instrument. Since the TB820D
cannot detect whether the sample water supply is stopped or not, periodical check of the sample
line is needed.
6.4
Automatic wash/Automatic calibration
Operation
When the sampling system is correctly installed, and Automatic wash/Automatic calibration
is properly setup on the converter setting, Automatic wash/Automatic calibration starts at a
scheduled time during a regular operation in a measurement mode. For the details about the
schedule, see 4.5 Auto wash/calibration setting.
You can check when the next auto sequence starts on the sensor detail screen. If the scheduled
auto sequence date has already passed, Auto wash/Auto calibration does not operate.
For details about the converter displays or mA displays during Automatic wash/Automatic
calibration operation, see FLXA402T Operation of Converter
. During
Automatic wash/Automatic calibration, you cannot turn on the maintenance mode. You cannot
enter the calibration menu or the setting menu, either.
6.5
Stop and restarting operation
To stop an operation, stop supplying power. If the instrument will be out of operation for a long
period, stop supplying sample and wash the Flow cell and keep it empty or keep zero water
running. When restarting operation, perform necessary inspection and maintenance referring to
the chapter 8. Maintenance.