<5. FLXA402T sensor menu>
5-5
IM 12E01B30-02EN 2nd Edition : June 09 , 2021-00
n
TB detector
Displays Serial number (Serial No.) of a connected Turbidity detection unit, software version
(Software Rev.), Model code (Model code), internal serial number of CPT board, LED board and
analog board (Internal SN.(CPT), Internal SN.(LED), Internal SN.(Analog) ).
n
Converter log, Sensor log
Same display as on the converter “Detail”.
See FLXA402T Operation of Converter
5.4
Reset op. time
To reset wellness data of each sensor
Sensor menu > Reset op. time > Sensor wellness , tap “Reset data” > Reset data
You can reset each data from the dialog of Sensor wellness. If you reset, the operating days are
reset to “0” but the replacement date is updated.
When you have a password, the password verification appears. Upon successful confirmation,
Reset confirmation screen appears.
For further information on the password setting, see FLXA402T Operation of Converter
5.5 Setting
Sensor menu > Setting .
When you have a password, the password verification appears. Upon successful confirmation,
Sensor Setting screen appears.
For further information on the password setting, see FLXA402T Operation of Converter
If you go to Sensor setting, the mA output becomes HOLD when automatic hold function is
enabled, and contact output keeps the current status.