![YOKOGAWA TA320 Скачать руководство пользователя страница 15](http://html.mh-extra.com/html/yokogawa/ta320/ta320_user-manual_911358015.webp)
1-5
IM 704210-01E
Functions
1
1.4 Computation
Readout Function
≡
For operating procedure, refer to 7.1 and 7.5
≡
If histogram or time variation display format is selected (see 1.6), two markers (X1 and X2) will
be displayed. This enables reading out of the coordinates of each cross-point of the markers and
the histogram (or time variation). In addition, selecting a calculation area (see the following
page) enables statistical computation for the area enclosed by the markers.
Histogram display format
The meaning of each value displayed at the bottom of the screen is given below.
X1: X-axis coordinate of X1 marker (measured value)
Y1: Y-axis coordinate of X1 marker (frequency of X1)
X2: X-axis coordinate of X2 marker (measured value)
Y2: Y-axis coordinate of X2 marker (frequency of X2)
N:
• WINDOW selected as calculation area: Total number of samples within the window
• MARKER selected as calculation area: Number of samples present within the area enclosed
by X1 and X2 markers
X1 marker
X2 Marker
Window
Time variation display format
The meaning of each value displayed at the bottom of the screen is given below.
X1: X-axis coordinate of X1 marker (time stamp)
X2: X-axis coordinate of X2 marker (time stamp)
∆
X: X2 - X1
Y1: Y-axis coordinate of X1 marker (measured value)
Y2: Y-axis coordinate of X2 marker (measured value)
∆
Y: |Y2 – Y1|
X1 marker
X2 Marker