8-26
IM 704420-01E
8.7 Commands
8.7.8 MEMory Group
The commands in this group deal with the external transmission of the measured data. The raw data that
has been measured is externally transmitted in ASCII or binary format. For a detailed description of this
function, see the next page.
;
:MEMor y
:
DATaselect
<Space>
TSTamp
MEASuredata
FREQuency
?
FORMat
<Space>
ASCii
BINar y
?
BYTeorder
<Space>
LSBFirst
MSBFirst
?
STARt
?
END
?
SIZE
?
SEND
?
?
: M E M o r y ?
: M E M o r y ?
: M E M o r y ?
: M E M o r y ?
: M E M o r y ?
Function
Queries all settings related the external
transmission of the measured data.
Syntax
:MEMory?
Example
:MEMORY?
→
:MEMORY:DATASELECT FREQUENCY;
FORMAT ASCII;BYTEORDER LSBFIRST
: M E M o r y : B Y T e o r d e r
: M E M o r y : B Y T e o r d e r
: M E M o r y : B Y T e o r d e r
: M E M o r y : B Y T e o r d e r
: M E M o r y : B Y T e o r d e r
Function
Sets the transmission order of binary data or
queries the current setting.
Syntax
:MEMory:BYTeorder {LSBFirst|MSBFirst}
:MEMory:BYTeorder?
Example
:MEMORY:BYTEORDER LSBFIRST
:MEMORY:BYTEORDER?
→
:MEMORY:
BYTEORDER LSBFIRST
: M E M o r y : D A T a s e l e c t
: M E M o r y : D A T a s e l e c t
: M E M o r y : D A T a s e l e c t
: M E M o r y : D A T a s e l e c t
: M E M o r y : D A T a s e l e c t
Function
Sets the data to be transmitted or queries the
current setting.
Syntax
:MEMory:DATaselect {TSTamp|MEASuredata|
FREQuency}
:MEMory:DATaselect?
Example
:MEMORY:DATASELECT FREQUENCY
:MEMORY:DATASELECT?
→
:MEMORY:
DATASELECT FREQUENCY
Example
The transmitted data varies depending on the
measurement function.
• When the measurement function is set to
“
DtoC
” or “
3T
”:
MEASuredata (measured time data) and
FREQuency (histogram frequency data)
• When the measurement function is set to
“
WOBBle
” (BI-PHASE measurement function
option, see the TA120E Digital Jitter Meter
Option Function User’s Manual (IM704420-
51E)):
TSTamp (measured time stamp) and
MEASuredata (measured time data)