IM 704420-01E
9.4
Performing a Self-Test
Keys
POWER
KEY
LOCK
MEAS
SET
AVE
/T
0
2
4
4
8
6
12
8
16
10
%
deg
%
20
180
us
ms
ns
MEAS
x1
3TCD
xN
3TCD
xN
BI-PHASE
SPEED
DELAY
GATE TIME
FUNC
GATE
LOCAL
DATA
SCALE
PRESET
PHASE
10
20
TRIG
CLOCK
PHASE
AUTO MAN
GP-IB
REMOTE
EQLZ
LEVEL
PLL
SHIFT
ARM
INH
DtoC
DVD
0.1s
0.5s
MAN
UTILITY
OPTION
10
5
σ
σ
Procedure
Removing the signal cable
1.
Remove all cables that are connected to the TA120E (input and output cables)
excluding the power cable and GP-IB cable. Make sure that the TA120E is not
in the remote mode at this point.
For the procedure in switching from remote mode to local mode, see section 8.2.
Setting the TA in the maintenance mode
2.
Check that the power switch is turned OFF.
3.
While pressing the SCALE key, turn ON the power switch. Hold the SCALE key
down for approximately 3 seconds. After displaying [TA120E
→
704420
→
V-
x.xx(version display)
→
tESt
→
PASS, the characters [tEST] appear on the
display. The TA120E enters the start condition in the maintenance mode.
Testing the keys and rotary knob
• Performing the test
4.
Press the FUNC key. The characters [KEY] appear on the display.
5.
Press a key. Confirm that the character corresponding to the pressed key is
shown on the display. In addition, confirm that a character indicating the
rotation direction is shown on the display when the rotary knob is turned.
• Returning to the start condition in the maintenance mode.
6.
Press the < key twice. The characters [tEST] appear on the display and the
TA120E enters the start condition in the maintenance mode.