7-2
IM 701240-01E
Conditions When Waveform Acquisition Cannot Be Started or Stopped
• When the SL1400 is being controlled remotely via communications.
• While printing is in progress or during auto setup.
Waveform Acquisition Mode at Power ON
You can select whether to start the waveform acquisition at power on. For example,
select ON if you want to start the waveform acquisition when power supply recovers after
a power failure. The default setting is OFF.
ON:
Starts waveform acquisition at power on.
OFF: Does not start the waveform acquisition at power on.
Response Time of the START/STOP Key
You can set whether the SL1400 responds immediately when START/STOP is pressed
or when the key is held down for at least 1 second.
Quick:
Immediate response.
>1sec:
Responds when the key is held down for at least 1 second.
Note
•
If you start the waveform acquisition after changing the waveform acquisition conditions
indicated below, the past data that is stored in the acquisition memory is cleared.
•
Setup conditions on the ACQ menu
•
Setup conditions of the module
•
Input coupling of the 701261 (UNIVERSAL), 701262 (UNIVERSAL (AAF)), or 701265
(TEMP/HPV) (when a change is made between TC and some other setting) and the
thermocouple type setting (when measuring temperature)
•
Input coupling of the 701275 (ACCL/VOLT) (when a change is made between ACCL and
some other setting) and the sensitivity setting (when measuring acceleration)
•
FV and offset settings of the 701280 (FREQ)
•
Trigger setup conditions
•
A snapshot function that keeps the current displayed waveform on the screen is also
available. You can update the display without stopping waveform acquisition (see section
8.7).
•
For a description of the behavior when the Start Mode (Power On) is ON or OFF, see
appendix 8.
7.1 Starting/Stopping Waveform Acquisition