Waveform Acquisition
7.1
Starting/Stopping Waveform Acquisition
Procedure
Starting/Stopping Waveform Acquisition
Press
START/STOP
. Waveform acquisition starts or stops.
Waveform acquisition is in progress when the RECORD indicator above the START/
STOP key is illuminated.
Setting the Waveform Acquisition Mode at Power ON
1.
Press
MENU
.
2.
Press the
Next 1/2
soft key.
3.
Press the
Others
soft key.
4.
Press the
Start Mode(Power On)
soft key to select ON or OFF.
Setting the Start/Stop Response Time
5.
Press the
START/STOP Response Time
soft key to select Quick or >1sec.
Explanation
Waveform Acquisition and Indicator Display
• Waveform acquisition is in progress when the RECORD indicator above the START/
STOP key is illuminated. The word “Running” is displayed at the lower left of the
screen.
• If the RECORD indicator above START/STOP is not illuminated, waveform acquisition
is stopped. The word “Stopped” is displayed at the lower left of the screen.
Operation When the Acquisition Mode Is Set to Averaging Mode
• Averaging is stopped when waveform acquisition is stopped.
• Averaging starts again when waveform acquisition is restarted.
START/STOP Operation during Accumulation
• Accumulation is suspended when acquisition is stopped.
• It is resumed when waveform acquisition is restarted.
Chapter 7