10-4
IM DL950-02EN
Setting Measurement Items (Item Setup)
4.
Tap
Item Setup
, or press the soft key. The following screen appears.
Select the measurement items that you want to use.
X1: Time value at Cursor1
X2: Time value at Cursor2
ΔX: Difference between the time values of Cursor1 and Cursor2
1/ΔX: Reciprocal of the difference between the time values of Cursor1
and Cursor2
Y1: Vertical value at the intersection of Cursor1 and the waveform
Y2: Vertical value at the intersection of Cursor2 and the waveform
ΔY: Difference in the vertical values at the intersection of
Cursor1/Cursor2 and the waveform
Cursor Value Read Mode
4.
Tap
Other Setup
, or press the soft key.
5.
Tap
Cursor Read Mode
, or press the soft key to set the cursor value read mode.
The two modes available for reading Y1 and Y2 measurements are Display (display data) or
ACQ (acquisition data).
Display: The screen is divided into 1001 parts, and the maximum and minimum values
displayed on those rasters are read.
ACQ:
Measurement data acquired at the sample rate of each waveform is read. When all
traces are selected, measurement data acquired at the sample rate calculated from
T/div and record length is read.
10.2 Measuring with Vertical Cursors