11-56
IM 701210-06E
Output Signal
NOGO OUT (Negative Logic)
When the determination result is “NO-GO” (fail), the output signal level changes from
high level to low level temporarily.
GO OUT (Negative Logic)
When the determination result is “GO” (pass), the output signal level changes from high
level to low level temporarily.
Signal Output Circuit
GO OUT
NO-GO OUT
5 V
120
Ω
4.7 k
Ω
100 pF
GO/NO-GO I/O Timing
START IN
Waveform acquisition
Waveform acquisition
Decision
Result
Decision
Result
NOGO OUT
GO OUT
100 ms or more
50 ms
or more
0 ms
or more
Remains at low output until the next determination
can be accepted.
The time is extended while outputting to the built-in
printer or internal storage medium.
When Remote Start is OFF, START IN is not used.
GO OUT and NO-GO OUT are output using the
internal determination timing.
GO/NO-GO Cable (366973, Sold Separately)
• Use this cable only for the GO/NO-GO determination on the DL750/DL750P.
• See the following figure for the connection to external devices.
1
6
Approx. 1.5 m
RJ-11 connector
To the DL750/DL750P
Wires are separated.
Modify them to suit your needs.
Pin No.
Signal Name
Logic
2
3
4
5
START IN
GO OUT
NOGO OUT
Negative
Negative
Negative
GND
Color
Yellow
White
Green
Blue
11.10 Using the GO/NO-GO Determination I/O Function