11-47
IM 701210-06E
Waveform Analysis/Search
11
Explanation
GO/NO-GO is determined by creating a zone based on a reference waveform and
checking whether or not the waveform has left or entered the zone.
Selecting the Reference Waveform: Edit Zone
Select the reference waveform used to create the determination zone. You can select
any of the following waveforms, that are displayed as a trace, as the reference waveform.
CH1 to CH16, DSP1 to DSP6 (optional), Math1 to Math8
Creating the Determination Zone
You can specify up to 6 determination zones. The selectable range is as follows:
• Vertical range:
±
10 divisions from the reference waveform
• Horizontal range:
±
5 divisions from the screen center
You can select input signal waveforms (CH1 to CH16), computed waveforms (Math1 to
Math8), and DSP channel waveforms (DSP1 to DSP6, optional)) for the waveforms that are
to be determined in the determination zones that are registered in Zone1 through Zone6.
The zones that are enabled through Setup are displayed on the screen.
Note
Match the input type of the target waveform to the reference waveform that was used when
the zone was created.
For example, if the reference waveform was created from a “voltage input waveform,” set the
target waveform to “voltage input waveform.” Correct determination cannot be made if the
target waveform is set to “strain” or “temperature.”
Setting the Determination Conditions: Setup
• Selecting the Target Waveform: Trace
Select the waveform from CH1 to CH16 and Math waveforms.
• Setting the Determination Mode
IN:
GO condition when all the applicable waveforms enter the determination zone.
NO-GO if any of the waveforms exists the determination zone.
OUT:
GO condition when all the applicable waveforms are outside the determination
zone. NO-GO if any of the waveforms enters the determination zone.
OFF:
Does not perform GO/NO-GO determination.
• Zone Number: ZoneNo.
Select the zone number from Zone1 to Zone6.
• Setting the Logic
AND:
Executes the action when all conditions (1 to 16 types) are met.
OR:
Executes the Action when any one of the conditions (1 to 16 types) is met.
• Action Condition: ActCondition
Always:
Always executes the action.
Failure:
Execute the action when the GO condition is not met.
Success:
Execute the action when the GO condition is met.
When Always is selected, the operation specified by Action is executed each time the
trigger occurs. This function is useful such as when you wish to print the screen
image data to a printer each time the trigger occurs.
• Setting the Number of Actions: Sequence
Single:
Performs the action once.
Continue: Repeats the action up to acquisition count specified by ACQ Count. (If the
count is set to Infinite, the action is repeated until the waveform acquisition
is stopped.)
11.8 Performing GO/NO-GO Determination Using Zones