5-8
IM DL350-01EN
Trigger mode (Trigger Mode)
This is a setting in scope mode.
The trigger mode determines the conditions for updating the displayed waveforms. There are four trigger modes.
Auto Mode (Auto)
If the trigger conditions are met before an approximately 50 ms timeout, the instrument updates the displayed
waveforms on each trigger occurrence. If not, the instrument automatically updates the displayed waveforms. If
simple trigger is in use and the trigger source is set to Time, the instrument operates in Normal mode even when
Auto mode is specified.
If the time axis is set to a value that would cause the display to switch to roll mode, roll mode display will be
enabled.
Normal Mode (Normal)
The instrument updates the waveform display only when the trigger conditions are met. If no triggers occur, the
display is not updated. If you want to view waveforms that the instrument cannot trigger on, or if you want to
check the ground level, use Auto mode.
Single Mode (Single)
When the trigger conditions are met, the instrument updates the displayed waveform once and stops waveform
acquisition.
If the time axis is set to a value that would cause the display to switch to roll mode, roll mode display will be
enabled. When the instrument triggers, it begins recording data. When data has been acquired up to the amount
specified by the set record length, the waveform display stops.
Instant Start Mode (On Start)
Regardless of the trigger settings, when you press the START key, the instrument updates the displayed
waveforms once and stops signal acquisition.
If the time axis is set to a value that would cause the display to switch to roll mode, roll mode display will be
enabled. When data has been acquired up to the amount specified by the set record length, the waveform
display stops.
• The trigger mode setting applies to all trigger types.
• When waveforms are being acquired, the trigger condition appears in the center of the bottom of the
screen.
5 Waveform Acquisition