4-65
IM DL350-01EN
Position Information (GPS)
For position information, set whether to turn the display on or off, the items to be set, measurement range, label,
display range, display group, and the like.
Installing a GPS unit (720940/B8093YA, an accessory sold separately) enables the following items to be
measured.
• Latitude (Latitude) [°]
• Longitude (Longitude) [°]
• Altitude (Altitude) [m]
• Velocity (Velocity) [km/h]
• Direction (Direction) [°] with North assumed to be 0°
• Status (Status)
• Bit1: 3D FIX. Set to 1 when 3D position is determined.
• Bit2: PPS (Pulse Per Second). Set to 1 when time pulses synchronized to the GPS time are obtained.
Display ON/OFF (GPS)
Set whether to show or hide position information.
• ON: Position information is shown.
• OFF: Position information is hidden.
To acquire position information, you need to set
All Items Setup (All Items Setup)
On a single screen, you can set the measurement range, label, display range, position, magnification, and the
like for each position information entry
Items to Be Set (Item)
Select the items to be set for each position information entry, such as the display on/off state, measurement
range, label, and display range.
• Latitude: Latitude [°]
• Longitude: Longitude [°]
• Altitude: Altitude [m]
• Velocity: Velocity [km/h]
• Direction: Direction [°] with North assumed to be 0°
• Status: 3D FIX, PPS status
Display ON/OFF State of Each Position Information Entry
Set whether to show or hide each position information entry.
• ON: Position information is shown. If the waveform display (described earlier) is also set to ON, the position
information that is set to ON here will be displayed on the screen.
• OFF: Position information is hidden.
Measurement Range (Range)
Set the measurement range for the altitude (Altitude) and velocity (Velocity).
• Altitude (Altitude)
3276.7m: The measurement range is set to −3276.7 to 3276.7 m. The measurement resolution is set to 0.1 m.
32767m: The measurement range is set to –32767 to 32767 m. The measurement resolution is set to 1 m.
• Velocity (Velocity)
655.35km/h: The measurement range is set to 0 to 655.35 km. The measurement resolution is set to 0.01 km/h.
6553.5km/h: The measurement range is set to 0 to 6553.5 km. The measurement resolution is set to 0.1 km/h.
4 Vertical Axis