4-48
IM DL350-01EN
• Terminator (Terminator)
ON: The 120 Ω terminator between CAN_H and CAN_L on the CAN bus line is turned on.
OFF: The terminator between CAN_H and CAN_L on the CAN bus line is turned off.
CAN I/F
CAN TRANSCEIVER
CAN_H
60Ω
7
2
3
6
60Ω
CAN_H
CAN_L
CAN_L
GND
GND
0.1 μF
CAN Data Extraction Conditions (All SubChannel Setup)
Set the extraction conditions for the CAN data that is in the data field. You can configure the settings for each
sub channel. There are up to 60 sub channels for each port.
When the data field contains two units of data
bit 64 56 48 40 32 24 16 8 0
CH3_2
CH3_1
• Start Bit = 0, Bit Count = 16bit → 2-byte integer (CH3_1)
• Start Bit = 32, Bit Count = 32bit → 4-byte real (CH3_2)
• Input (Input)
ON: The data is monitored.
OFF: The data is not monitored or displayed.
• Labels (Label)
• Message Format (Msg Fmt)
Select the message format of the collected data frames.
STD: Standard format
XTD: Extended format
• ID (Hex)
Set the message ID of the collected data frames.
Standard format (11 bits): 0x000 to 0x7ff
Extended format (29 bits): 0x00000000 to 0x1fffffff
• Byte Count
Set the method for extracting the data area of the collected data frames.
Auto: All data is extracted. Normally use this option.
1 to 8: The specified number of bytes of data are extracted. Data is extracted starting from the start of the
data. Byte Count is enabled when big endian byte order is in use. When little endian byte order is in
use, Byte Count is always set to Auto.
4 Vertical Axis