9-5
IM DL350-01EN
Marker Cursors (Marker) - T-Y waveforms
Four markers are displayed on the selected waveform. You can measure the level at each marker, the amount of
time from the trigger position to each marker, and the level and time differences between markers.
Marker1
X
Y
+
Y
Marker2
Marker3
Marker4
Marker Settings (Marker Setup)
Measurement Source Waveform (Trace)
Set the measurement source waveform to one of the waveforms below.
• OFF: Not use
• CH1 to CH4
1
, 16chVOLT
2
, 16chTEMP/VOLT
2
, CAN
3
, LIN
3
, SENT
3
, Math1 to Math2
1 You can select the channel of an installed module. On a 4-CH module, select sub channel 1 or 2. You
cannot select the input channel of a logic module.
2 When a 16-CH voltage input module or 16-CH temperature/voltage input module is installed. After you
select 16chVOLT or 16chTEMP/VOLT, select a sub channel.
3 On a model with the /VE option when a CAN bus monitor, CAN & LIN bus, or SENT monitor module is
installed. After you select CAN, LIN, or SENT, select a sub channel. This cannot be selected on a CAN bus
monitor or CAN & LIN bus monitor module if the data type (Value Type) is set to Logic. On a SENT monitor
module, S&C and Error Trigger sub channels cannot be selected.
Marker Shape (Marker Form)
Set the shape of the displayed marker to one of the options below.
• Mark: A dot
• Line: A crosshair
9 Cursor Measurement