<6. Converter Operating Functions>
6-5
IM 12F4A1-01E
6.5
Default Status When Mode Change
Table 6.4 shows default status of cell motor, metering pumps, and solenoid valves (SV1 to SV6)
when mode change.
Table 6.4
Default Status of Pumps, Solenoid Valves, etc. When Mode Change
Equipment Name/
Symbol
<MEAS.> Mode
<MAINT.> Mode
<PROGRAM.1>
Mode
<PROGRAM.2>
Mode
RC400G-1, 2 RC400G-3 RC400G-1, 2 RC400G-3
Cell motor
Metering pump
Air pump *1
ON
ON
ON/OFF
ON
ON
ON/OFF
ON
ON
ON/OFF
ON
ON
ON/OFF
*2
*2
SV1 (FILTER 1)
SV2 (WASH)
SV3 (LINE 1)
SV4 (LINE 2)
SV5 (FILTER 2)
SV6 (SAMPLE/ZERO)
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
ON
ON
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
ON
OFF
OFF
OFF
*1:
The air pump is turned OFF only when "oFF" is selected with function code "d." in <PROGRAM.1> mode.
*2:
In <PROGRAM.1> and <PROGRAM.2> modes, the setting data is kept in last mode before mode change.
14th Edition : Jul. 15, 2020