<3. Installation, Tubing and Wiring>
3-6
IM 12F4A1-01E
V6
PR
V3
V4
COM
NO
NC
SV6
F11
AP
S1
SV1
SV3
V1
V2
V5
SV4
V7
SV2
SV5
PU
CELL
CON
F2
F1
V8
COM
V12
V11
ø6 x ø4
polyethylene tube
ø8 x ø6
polyethylene tube
ø22 x ø15
braided wire reinforced
soft PVC tube
ø33 x ø25
braided wire reinforced
soft PVC tube
VP16 pipe
VP40 pipe
* option
Measuring sample (VP16)
Tap water (VP16)
Standard solution (Rc1/4)
Reagent (Rc1/4)
Air (Rc1/4)
Drain (VP40)
*Air pump
Vat
Strainer
Pressure reducing valve
Ball valve
Ball
valve
Activated
charcoal
filter
Converter
Measuring
tank
Metering
pump
Sand filter
Sand filter
3-way solenoid
valve
Solenoid valve
Tubing materials
F1_4_6E.ai
Figure 3.3.3
Tubing Diagram of RC400G-3/AZC
3.2.1
Sample Water Tubing
This is the tubing for taking sample water and feeding it to the analyzer.
The connection is a rigid PVC tube of nominal dia. 16 (O.D. 22 mm). Implement tubing using
proper joints, i.e. unions and flanges etc., which fit the connection. Note: Perform tubing so that
sample water pressure at the connection falls within the range of 20 to 500 kPa.
Sample water flow rates to the analyzer in operation are as follows:
In RC400G-1:
1 to 4 l/min
In RC400G-2:
5 to 10 l/min
In RC400G-3:
10 to 20 l/min
3.2.2
Reagent Tubing
Residual chlorine content is determined by sample water mixed with reagent. The reagent tubing
supplies reagent to the measuring cell of the analyzer.
Connect the outlet of the reagent tank to the reagent inlet of the analyzer using Rc1/4 joint and
O.D. 6 mm x I.D. 4 mm polyethylene tube.
There is a reagent level limit in the tank.
Check that the position of the reagent tank is in accordance with Figure 2.4 RC401G Reagent
Tank.
14th Edition : Jul. 15, 2020