17-12
IM 253710-01E
17.13 General Specifications
Item
Specification
Standard operating
Ambient temperature : 23
±
3
°
C
conditions
Ambient humidity : 50
±
10% RH
Power voltage and frequency : Less than 5% of the rated voltage/frequency
After performing zero level compensation after the warm-up time (at least 30 min.) has passed.
Operating conditions
Temperature : 5 to 40
°
C
Humidity : 20 to 85% RH when not using the printer, 35 to 80% RH when using the printer.
Operating altitude
2000 m or less
Storage environment
Temperature : –25 to 60
°
C
Humidity : 20 to 80% RH (no condensation)
Rated supply voltage
100 to 120 VAC, 200 to 240 VAC
Permitted supply voltage
90 to 132 VAC, 180 to 264 VAC
range
Rated supply voltage
50/60 Hz
frequency
Permitted supply voltage
48 to 63 Hz
frequency range
Power consumption
200 VA (when using the printer)
Insulation resistance
Between power plug and case : 50 M
Ω
or more at 500 VDC
Withstanding voltage
Between power plug and case : 1500 VAC (50/60 Hz) for one minute
External dimensions
Approx. 426 mm(W)
×
177 mm(H)
×
450 mm(D)
(For details, see page
(when the printer cover is closed, excludes the handle and other projections.)
17-14.)
Weight
Main unit only (253710) : Approx. 10 kg
Main unit when four power measurement modules (253752) are installed : Approx. 15 kg
Cooling method
Forced air cooling
Installation position
Horizontal (except the stand can be used). Vertical position and stacking not allowed.
Battery backup
Setup parameters and internal time clock are backed up with lithium batteries.
Fuse
Location
Max. rated voltage
Max. rated current
Type
Standard
Power supply
250 V
6.3 A
Time lag
UL/VDE certified
SCSI board*
250 V
800 mA
Time lag
VDE/SEMKO certified
*
The user cannot replace this fuse, because it is located inside the case. If you believe the
fuse is blown, contact your dealer.