9-25
IM 253710-01E
Waveform Display
9
Zooming in on the vector
(This is applicable to products (PZ4000) with firmware version 2.01 or later.)
7.
Press the [U Mag] or [I Mag] soft key to set the jog shuttle control to [U Mag], [I
Mag], or both [U Mag] and [I Mag].
8.
Turn the jog shuttle to set the zoom factor.
Explanation
Vector display is possible during harmonic measurement mode.
Selecting the display format
Select [Vector] to set the vector display. The phase difference and the amplitude
(rms) relationships between the fundamental signal U(1) and the I(1) of the element
wired using wiring method A can be displayed as vectors. The vector of each input
signal is displayed with the positive vertical axis representing zero degrees.
Turning ON/OFF the numerical data display
Select whether or not to display the numerical data. The values of the amplitude and
the phase difference between the signals can be displayed along with the vectors.
For information pertaining to the display format of the phase difference, see section
10.6, “Selecting the Display Format of the Phase Difference.”
• ON : Displays the numerical data.
• OFF : Does not display the numerical data.
Setting the zoom factor of the vector
The size of the vectors can be changed. This is applicable to products (PZ4000) with
firmware version 2.01 or later.
• The zoom factor can be set in the range from 0.100 to 100.000.
• You can specify separate zoom factors for the fundamental waves U(1) and I(1).
9.9 Displaying the Vectors of the Harmonics