7-14
IM 253710-01E
Explanation
The trigger position matches the trigger point when the trigger delay is set to 0 s. This
function allows you to display the waveform that is acquired the specified amount of time
after the trigger point (delay time).
Setting the trigger delay
Set the trigger delay in the range from 0.0 to 1000000.0
µ
s (with 0.5
µ
s resolution).
Delay time
Trigger point
Trigger position
When a trigger delay is set
Note
The trigger delay setting is retained even if the observation time is changed.
7.6 Setting the Trigger Delay