Setting the Time Axis
6
6.3
Selecting the Time Base
«For a functional description, see section 1.3.»
Keys
• The
mark indicates the keys that are used for the operation.
• To exit the menu during operation, press the ESC key.
CONFIGURATION
ELEMENTS
VOLTAGE
CURRENT
FILTER
1
FILTER
FILTER
FILTER
2
3
4
RES
ET SELECT
CH
8
CH
6
CH
4
CH
2
CH
7
CH
5
CH
3
CH
1
OBSERVATION TIME
TRIG’ D
ACQ
REMOTE
MENU
SHIFT
NULL
CAL
COPY
LOCAL
HELP
MISC
FILE
ZOOM
CURSOR
MATH
DISPLAY
SETUP INPUT MEASURE TRIGGER
ABORT
E S C
SINGLE
START
START/STOP
Procedure
1.
Press the SHIFT+TRIGGER(ACQ) key to display the Acq setting menu.
2.
Press the [Time Base] soft key to select [Int] or [Ext].
Explanation
With the default setting of the instrument, the data sampling timing is controlled by the
clock signal that is generated by the internal time base circuit. This setting can be
changed so that the timing is controlled by an external clock signal.
Selecting the time base
Select the time base from the following choices.
• Int : The internal clock signal becomes the time base. The observation time that
was set in section 6.1 becomes effective.
• Ext : The clock signal applied to the external clock input connector becomes the
time base. The observation time that was set in section 6.1 becomes ineffective.