![YOKOGAWA PH72 Скачать руководство пользователя страница 30](http://html.mh-extra.com/html/yokogawa/ph72/ph72_user-manual_911303030.webp)
IM 12B03D02-01E
5-1
5
.
Keypad and Display Functions
5. Keypad and Display Functions
There are eight membrane keys on the keypad of the Model PH72 Personal pH
Meter. The following key functions are provided.
• Displaying a pH (or mV) value and temperature
• Displaying a pH (or mV) value, date and time
• Holding a pH (or mV) value and temperature
• Storing data — pH (or mV) value and associated information
• Function mode
• Calibration
POWER
Auto Power Off
HOLD
F/ENT
MEAS
DATA
F050001.ai
CAL
Turns power on or off.
Automatic Power Off
function (default: 20 min).
Moves to calibration
mode.
Moves to measurement
mode or switches
measurement display
panels. Also exits from the
current function and returns
to measurement mode.
Changes setting or
value.
Moves to function
mode or confirms data
entry.
Holds the currently
measured value.
Stores the currently
measured value in
memory.
5.1 Keypad Functions
POWER
: Power On/Off key
Pressing and holding this key for at least one second when nothing is displayed on
the LCD, will turn the meter on. The meter will be turned off by pressing and holding
this key for at least two seconds when the meter is on. If no keys are pressed for a
preset time, the meter turns off power automatically (refer to Section 5.3 (8), “Set Auto
Power Off time (A.oFF) panel”).
CAL
: Calibration key
When pressed during measurement,
mark turns on and the meter moves to
calibration mode. To return to measurement mode, press
CAL
or
MEAS
key.
HOLD
: HOLD key
When pressed during measurement,
mark turns on and the currently
displayed measured pH (or mV) value and temperature are held. Pressing
HOLD
or
MEAS
key will turn
mark off and return the meter to measurement mode.