<3. Installation>
3-3
IM 11M10B01-01E
3.2.2
Installing Panel Mount Type
(1) Attach a panel mount frame to the side of the OX400, insert it into the panel, and securely
screw it to the panel.
(2) Provide a distance of 100 mm or more behind the OX400 in order to not block the outlet of
the cooling fan on the rear panel.
(3) The air inlet of the cooling fan is located on the bottom panel of the OX400. Be sure to
provide a distance of 25 mm or more from the bottom panel in order to not block the bottom
panel of the OX400.
135
+2 0
101.6
±0.3
240
±1
225
±1
19.3
330
20
10 to 20
6
6
132
101.6
260
238
Cooling fan
Unit: mm
Do not block
the air inlet.
Air is discharged
from the cooling
fan
Distance of 100 or more
Seal
Frame
Frame
Screw
4xM5 screw
Panel Cutout
Mount
3. Make sure the panel is at least 2 mm thick.
2. Maintain at least 100 mm of free space around the measuring instrument in order to ensure
adequate ventilation.
1. Make sure the bottom supports do not block the ventilation outlet on the bottom panel of
the measuring instrument.
Notes on mounting
Distance of 25 or more
Figure 3.2
Notes on Mounting Panel Mount Type
8th Edition : Jan. 10, 2019-00