7-9
IM 04P01B01-01E
Setup Operations for Changing/Adding Functions (Basic Setting Mode)
1
2
3
4
5
6
7
8
9
10
11
12
App
Index
7.6 Recording by Compensating for the Pen Offset
along the Time Axis (Pen Model)
This section explains the details of setting the compensation for the pen offset along the
time axis (pen model).
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
Changing the Settings
1.
Hold down the
MENU
key for 3 seconds to enter Setting mode.
2.
Hold down the
and
keys simultaneously for 3 seconds to enter Basic
Setting mode.
3.
Press the
key or
SHIFT
+
key to select
POC
and then press the key.
4.
Set each item and press the key.
Use the
key or
SHIFT
+
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the
ESC
key in the middle of the operation, the settings entered up to
then are cancelled, and the display returns to a higher level menu.
5.
When the
Setting complete
screen appears, do either of the following:
To correct the setting, press the key.
If you are done, press the
ESC
key.
Select On to record by compensating
for the pen offset along the time axis.
The new setting takes effect.
ESC/?
POC
Setting complete
POC=Off
Basic=POC
Applying the Changes and Returning to Operation Mode
Press the
key or
SHIFT
+
key to select
End
and then press the key. Press
the
key to select
Store
and then press the key. The changes are applied, and
the screen returns to Operation mode. If you select
Abort
and press the key, the
changes are discarded, and the screen returns to Operation mode. Press the
ESC
key to
return to the
Basic=
screen.
Explanation
POC (Pen Offset Compensation)
On: Enables pen offset compensation.
Off: Disables pen offset compensation.
Note
When pen offset compensation is enabled, all pens are aligned with the reference pen (pen that
is furthest to the back: pen 2 on the 2-pen model, pen 3 on the 3-pen model, and pen 4 on the
4-pen model) for recording. When recording is started, only the reference pen moves until the
offset is compensated. This is not a malfunction.
When pen offset compensation is enabled, an asterisk is printed by the time on the periodic
printout.