9.16 Setting the Bar Graph Display Mode
This section explains the details of setting the bar graph display mode.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
Changing the Settings
1.
Hold down the
MENU
key for 3 seconds to enter Setting mode.
2.
Hold down the
and
keys simultaneously for 3 seconds to enter Basic
Setting mode.
3.
Press the
key or
SHIFT
+
key to select
Math
and then press the key.
4.
Press the
key or
SHIFT
+
key to select
Bar graph
and then press the
key.
5.
Set each item and press the key.
Use the
key or
SHIFT
+
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the
ESC
key in the middle of the operation, the settings entered up to
then are cancelled, and the display returns to a higher level menu.
6.
When the
Setting complete
screen appears, do either of the following:
Press the key to set other channels.
If you are done, press the
ESC
key.
Set the channel range.
The new setting takes effect.
ESC/?
Select the display mode of the bar
graph.
First channel Last channel
0A-0A Channel
Setting complete
Basic=Math
Math=Bar graph
CH=0A-0A
Graph=Normal
Applying the Changes and Returning to Operation Mode
Press the
key or
SHIFT
+
key to select
End
and then press the key. Press
the
key to select
Store
and then press the key. The changes are applied, and
the screen returns to Operation mode. If you select
Abort
and press the key, the
changes are discarded, and the screen returns to Operation mode. Press the
ESC
key to
return to the
Basic=
screen.
Explanation
Graph
Normal: Sets the base point of the bar graph to the left or right of the recording span
whichever is smaller in value.
Center: Sets the base point of the bar graph to the 50% position of the recording span.
Содержание mR10000 436101
Страница 1: ...User s Manual Model 436101 436102 436103 436104 436106 µR10000 Recorder IM 04P01B01 01E 8th Edition ...
Страница 2: ......
Страница 14: ...Blank ...
Страница 18: ...Blank ...
Страница 58: ...Blank ...
Страница 72: ...Blank ...
Страница 92: ...Blank ...
Страница 220: ...Blank ...
Страница 268: ...Blank ...